CELL-DYN
®
3200 System Operator’s Manual
5-71
9140181H—October 2001
Section 5
Operating Instructions
Specimen Type
Figure 5.31 Specimen Type Screen
The
[SPECIMEN TYPE]
key is used to select the type of specimen that will be run.
(See Figure 5.31.) When the
[SPECIMEN TYPE]
key is pressed, the screen displays
a list of the QC files and the following soft key labels:
PATIENT
QC SPECIMEN
BACKGROUND
FRAGILE WBC
LATEX
RESISTANT RBC
RETURN
The function of each key is discussed below.
Selecting a Specimen Type
Most samples will be run in the Patient mode. If a FWBC or RRBC flag is
displayed on the
581
screen, the presence of fragile WBCs or resistant RBCs is
suspected. Depending on the flag, select either Fragile WBC or Resistant RBC as
the Specimen Type and re-run the sample.