CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 8.1 Needle Sensors Check/Adjustment
3 - 123
3
7. Press (
↑
N) key to move Needle to Home position (UP).
8. Ensure aspiration hole in needle is centered in waste cavity line and tip of needle is flush with
bottom of Wash Block.
NOTE:Needle must not extend below bottom of Wash Block nor recess more than 0.040” (1mm).
Figure 3-65
Limit (Bottom):
Tip 1/2" (~1 cm) Above Platen
Hole Centered on
Waste Drain Cavity
Home: Tip Flush with
Bottom of Wash Block or
Recessed Up to .040 in.
Wash Block
E LOADER WITH
BROWN WASH BLOCK
SL-274