CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 6.4 Mixing Contact Sensor Check/Adjustment
3 - 94
3
Figure 3-47
[NON-CE MARK] F Loader - Top of Mixing Paddle must be 0.040-0.60in.(1.0-1.5mm) below
bottom of Sensor (when Mixing Assembly is home,
).
.040 - .060 in
(1.0 - 1.5 mm)
SL-270
[CE MARK and NON-CE MARK]
F-Loader