![Abbott CELL-DYN 3000 Operator'S Manual Download Page 88](http://html1.mh-extra.com/html/abbott/cell-dyn-3000/cell-dyn-3000_operators-manual_450973088.webp)
CELL-DYN
3000 System Operator’s Manual
3-15
9140240E — May 1995
Chapter 3 Principles of Operation
MONO-POLY Histogram
The scatter information is plotted in a histogram format with the relative
number of cells on the Y axis and the mononuclear and
polymorphonuclear size distribution data on the X axis.
LYM-BASO-MONO Histogram
The scatter information is plotted in a histogram format with the relative
number of cells on the Y axis and the lymphocyte, basophil and
monocyte size distribution data on the X axis.
WBC Parameters
is generally displayed as depicted in
numeric and graphic data are automatically displayed on the
in the format selected. After the WBC scatter information
has been plotted and the cells have been classified into the five
subpopulations, the instrument determines the WBC count by
enumerating the pulses above the dynamic threshold in the 0
°
channel.
The algorithms then determine the WBC and the percent of cells in each
subpopulation.