2-10
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Installation Chapter 2
2.
Connect the red tube (aspiration waste fluid) connector to the red
tube attached to the lower port of the aspiration needle wash block.
See
3.
Connect the clear tube (aspiration rinse fluid) connector to the clear
tube attached to the upper port of the aspiration needle wash block.
4.
Connect the blue tube (vent waste fluid) connector to the blue tube
extending from the right side of the tower.
5.
Connect the white tube (vent rinse fluid) connector to the white
tube extending from the right side of the tower.
6.
Locate a fifth tube, the translucent silicone tube (sample
aspiration), extending from beside the open sample aspiration
probe wash block on the Analyzer module.
7.
Then, locate the clear sample aspiration tube attached to the top of
the aspiration needle and sticking out of the top of the tower.
8.
Connect these two tubes by inserting the end of the aspiration tube
leading from the top of the aspiration needle into the silicone
sample aspiration tube.
Tube Racks Setup
1.
Remove the 11 tube racks from the Accessory kit and inspect each
for damage.
2.
Remove the bar code labels provided for the tube racks and inspect
them for damage. Apply bar coded rack ID labels to the indented
area on the side of each tube rack between tube positions 1 and 2.
Be sure the labels are positioned in the indentations. Apply the bar
coded rack labels to the sides of the tube racks (in the area
indicated) by starting at the top and working downward. Label
MUST reside inside recessed area provided. It is suggested that the
racks be labeled 1–10 and the end rack be labeled 99.
3.
recessed area provided on the top of the rack.
4.
On one rack only, apply the end rack sensor label to the indented
area on the side of the rack, and apply two end rack visual indicator
labels to the indented areas on top of the rack. These labels will
identify this rack as the end rack.