10-70
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting
Chapter 10
Table 10.3
Sample-Related Fault Conditions (Continued)
Auto-Sampler consecutive data faults/BL
– – – – – Probable Cause – – – –
– – – – Corrective Action – – – –
1.
During sample processing,
1.
Correct the situation on the
four consecutive incomplete
Analyzer as follows:
aspiration or metering faults
were detected and the
•
Check the appropriate
Sample Loader halted.
tubing for a plug
or pinch.
•
Press [CLEAR ORIFICE]
to clear the aperture.
•
refer to the instructions
given in
2.
Press the [CLEAR FAULT]
key displayed on the
automatically resumes
processing.
3.
If the problem persists,
follow the instructions
given in the
chapter for trouble-
shooting
incomplete aspiration
messages.