CELL-DYN
3000 System Operator’s Manual
10-69
9140240E — May 1995
Chapter 10
Troubleshooting
Table 10.3
Sample-Related Fault Conditions (Continued)
Mixing Error/BL
MIXING ERR is displayed on the
result.
SAMPLING ERR is displayed to the right of the MCHC result.
MIXING ERR is printed on the graphics report. A mixing error is always
accompanied by a sampling error; therefore, SAMPLING ERR is printed
on all reports.
– – – – – Probable Cause – – – –
– – – – Corrective Action – – – –
1.
The blood sensors did
1.
Check the sample tube to
not detect a sufficient
be sure it contains a
amount of sample
sufficient quantity of during
aspiration.
blood.
2.
section of this chapter or
the
remove any obstructions.
3.
4.
5.
Call for technical assistance.