Troubleshooting Chapter 10
Table 10.2: General Fault Conditions
The following abbreviations are used in this table:
BL — Bulletin Line
SB — Status Box
Information for each message is listed in order from the most likely
cause of the message to the least likely cause of the message. Therefore,
always troubleshoot a problem in this order. If a problem cannot be
resolved by the corrective action indicated in this table, call for technical
assistance.
__________________________________________________________
Not Ready: See DIAG or Not Ready: See Special/SB
If a fault condition has occurred, the word FAULT on the Analyzer status
indicator panel is illuminated in red.
– – – – Probable Cause – – – – – – – – Corrective Action – – – –
1. A situation that prevents 1.
the READY state has been
detected. See the
[FAULT REPORT] followed
problem. Refer to the
whichever is indicated, for for appropriate table for more
information. corrective action.
2. A diagnostic test was run 2.
The instrument must be using
one of the
initialized when the
keys. is complete. Initialize the
Analyzer by pressing the
[INITIALIZATION] key on
3. A special protocols
3.
and perform the action
necessary to complete the
procedure.
4. System malfunction. 4.
Call for technical assistance.