10-46
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting
Chapter 10
Table 10.1:
Instrument Status Conditions
The following abbreviations are used in this table:
BL — Bulletin Line
SB — Status Box
__________________________________________________________
Ready/SB
– – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – -
The Analyzer is ready to process samples.
__________________________________________________________
Selecting open/closed mode/SB
– – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – –
The [CHANGE SAMPLER] key was pressed and the Analyzer is
changing to the selected mode of operation.
Resume processing when the READY message is displayed in the status
box.
__________________________________________________________
Entering standby/SB
– – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – –
The instrument has been idle for four hours and therefore is
automatically performing a cleaning cycle before entering the
STANDBY state. (Pressing the [DAILY SHUTDOWN] key also
initiates this message.)
When the cycle is complete, press [PRIME] or [RUN] to initiate the
priming cycle and return the instrument to the READY state. Resume
operation when the cycle is complete.
__________________________________________________________
Standby/SB
– – – – – – – – –– – – – – Explanation/Action– – – – – – – – – – – – – – -
The instrument has entered the STANDBY state. Press [PRIME] or
[RUN] to initiate the priming cycle and return the instrument to the
READY state. Resume operation when the cycle is complete.