CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting
Chapter 10
18.
Run the two diluent tubes and verify proper needle movement.
Check to be sure the background count is acceptable on the last
cycle.
NOTE: If either needle does not move up and down smoothly,
carefully check to be sure the locking screw is tight and that the
tubing attached to the needle moves freely. If any problems are
encountered, call for technical assistance. If the background
count is unacceptable, clean the needle as directed in
, and repeat the background count.
19.
Disconnect the sample aspiration tubing from the Analyzer.
20.
Replace the tower cover and check the tubing to ensure it is not
pinched or crimped. Replace the manual keyboard cover.
21.
Reconnect the sample aspiration tubing to the Analyzer.
CAUTION: Be careful not to crimp or bend the tubing when
reconnecting it. Do not use excessive force.
Aperture
It may be necessary to replace the RBC/PLT aperture if no other solution
to a related problem is found.
If you are replacing the aperture, verify that the correct aperture has been
received. RBC/PLT apertures are identified by R/P etched on the plate.
Use the following procedure to replace the aperture.
1.
2.
Remove the old aperture as directed in the
3.
Clean the new aperture and install it as directed in the aperture
cleaning procedure, and follow the remaining steps in that
procedure.
CAUTION: Changing the aperture may alter instrument
calibration. Therefore, confirm the calibration by running
commercial controls before processing samples. If necessary,
recalibrate the instrument as directed in
4.
Repeat the samples that were run when the problem was detected to
determine if it has been resolved. If the problem is not resolved,
call for technical assistance
at 1 (800) CELL DYN.