Chapter 6
Calibration
5.
Type any comments in the <COMMENTS> field. Press the Enter
key on the keyboard to save the entry and advance the cursor.
NOTE: Comments may be added to the Calibration Log only
after a calibration factor is changed or reentered.
6.
If desired, press [PRINT LOG] to obtain a printout of the
Calibration Log for documentation.
Calibration for Individual Parameters
If only certain individual parameters require calibration, continue with
this section:
1.
Press [QUIT AUTO-CAL] followed by [CONFIRM QUIT] to
exit Auto-Cal.
2.
Press [RETURN] twice to return to the
3.
Press [ENTER FACTOR]
ENTER WHOLE BLOOD FACTOR Screen
4.
Type any mean factors computed by Auto-Cal for those parameters
that require calibration in the appropriate fields. (Refer to the printout
made in
Calibration Factor Calculation
this procedure.) Move the cursor to the desired position, type the
factor and press the Enter key on the keyboard to save the factor and
advance the cursor.
5.
Press [RETURN].
6.
parameters that required calibration in the <COMMENTS> line.
Press the Enter key on the keyboard to save the entry and advance
the cursor.
NOTE: Manually entered factors are followed by the letter E in
parentheses to indicate “Enter Factor”.
7.
If desired, press [PRINT LOG] to obtain a printout of the
Calibration Log for documentation.
Completing Mode to Mode Calibration
1.
Press [RETURN] to return to the
2.
Press [MAIN] to exit the
3.
Retain all necessary documentation in the instrument logbook.