6-64
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
Table 6.6:
Mode to Mode Calibration Criteria Chart
Determining Which Parameters Need Calibration
1.
Validation Range
If the factor % difference is equal to or less than the following
values, CALIBRATION IS NOT REQUIRED for that parameter as
the value is within the Validation Range.
Results EQUAL TO OR LESS THAN VALIDATION RANGE:
WBC
<1
.75%
RBC
<1
.25%
HGB
<1
.25%
MCV
<1
.25%
PLT
<
3.50%
If the results for all parameters are less than the values given above:
Press [QUIT AUTO-CAL] followed by [CONFIRM QUIT].
Document the calibration as follows:
is available for comments only when
a calibration factor is changed or reentered. Therefore, document
the calibration in the <COMMENTS> line by accessing the log as
follows:
, press [ENTER
FACTOR].
Retype an existing Cal Factor. Press the Enter key on the keyboard
to save the entry and advance the cursor.
MODE TO MODE CALIBRATION CRITERIA
Validation Range
Cal Not Required
Calibration Range
Cal Required
Calibration Limit
Do Not Cal
WBC
<1.75%
>1.75% but
<
10%
>10%
RBC
<1.25%
>1.25% but
<
10%
>10%
HGB
<1.25%
>1.25% but
<
10%
>10%
MCV
<1.25%
>1.25% but
<
10%
>10%
PLT
<3.50%
>3.50% but
<
20%
>20%