Operating Instructions Chapter 5
3.
Open the well-mixed sample tube and immerse the open sample
aspiration probe in the sample.
4.
Press the touch plate located behind the probe to start the cycle.
The word “Busy” on the Analyzer status indicator panel is
illuminated in yellow. The Status Box on the
displays messages to indicate the various stages of the
cycle.
5.
Remove the tube when the beep sounds. The probe moves up
through the wash block to clean it.
6.
When the cycle is completed, the probe moves down, the word
“Ready” on the Analyzer status indicator panel is illuminated in
green and the results are displayed on the
7.
If automatic report printing has been specified, a report is printed
according to the options selected during Set Up. If this has not been
specified, a report may be printed by pressing [PRINT REPORT].
8.
Repeat this procedure for subsequent samples.
Closed Mode Analysis
The Closed Sampler Mode on Closed Sampler (CS) instruments
aspirates the sample from a closed collection tube that has been inserted
in the Closed Sampler module. CLOSED SAMPLER is displayed in the
upper right corner of the
selected. The probe is retracted when this mode is selected.
CAUTION: Damage to the instrument can occur if tubes are
incorrectly placed in the Closed Sampler module. Be certain to
invert sample tubes and place the stoppered end down into the
Closed Sampler module.
Closed Mode Procedure
1.
Be sure that the word “Ready” is illuminated on the Analyzer status
indicator panel and READY is displayed in the Status Box on the
2.
press [CHANGE
SAMPLER] to select the Closed mode.
3.
Invert the well-mixed sample and place the stoppered end down
into the Closed Sampler module. Push the end of the tube securely
into the tube retainer.