CELL-DYN
3000 System Operator’s Manual
4-7
9140240E — May 1995
Chapter 4
System Specifications
Power Specifications — CELL-DYN 3000CS
Table 4.6:
Power Specifications
Consumption
Analyzer:
900 Watts
Data Station:
300 Watts
Printer:
145 Watts
1345 Watts maximum (4590 BTU per hour)
Analyzer Input Requirements
Setting
Range
Frequency
100
90—110 VAC
50/60 Hz
120
110—130 VAC
50/60 Hz
220
200—240 VAC
50/60 Hz
240
220—250 VAC
50/60 Hz
Data Station Input Requirements
Setting
Range
Frequency
120
90—130 VAC
50/60 Hz
240
180—260 VAC
50/60 Hz
Printer Input Requirements
Setting
Frequency
120 VAC
50/60 Hz