7.

 

If the loop is not perpendicular to the shaft when extended from the sheath, completely retract the loop into the sheath and re-extend catheter. 

8.

 

The catheter should be passed from a peripheral vessel to the desired endocardiac position with the aid of fluoroscopy. Compatible navigation and 

visualization systems may be used in conjunction with fluoroscopy. 

9.

 

To adjust the curve of the distal tip on the uni-directional catheter, push or pull the thumb control located on the handle. To adjust the curve of the 

distal tip on the bi-directional catheter, use the actuator to deflect the catheter in either direction. 
NOTE: The bi-directional handle has an adjustable tension control knob that allows the operator to use the actuator and deflectable section in an 

unlocked state or adjust the tension to where the actuator and deflectable section are locked in place. The amount of friction increases as the 

knob is rotated clockwise until it reaches the fully plus (+) position. 

NOTE: The uni-directional handle has an adjustable tension control knob that allows the operator to use the actuator and deflectable section in an 

unlocked state or adjust the tension to where the actuator and deflectable section are locked in place. If necessary, the tension control knob may 
be rotated to increase or decrease the tension. 

10.

 

Prior to withdrawal, deflect catheter shaft to straight position. Re-insert the loop straightener into the hemostasis valve prior to removing the catheter 

from the introducer. 

11.

 

After use, dispose of product and packaging in accordance with hospital, administrative, and/or local government policy. 

Connection to Other Equipment 

The catheter may be connected to a commercially available EP recording system and navigation and visualization system using the connection cable. All 

systems must be patient isolated. For instructions regarding the use of these systems with the catheter, refer to the system’s instructions for use. 

Packaging and Shelf-Life 

The catheter packaging is designed to prevent crushing of the product, to minimize product exposure to the atmosphere, and to provide for aseptic product 

transfer. It is recommended that the product remain in the unopened package until time of use. Contents are sterile if the package is unopened and 

undamaged. Do not resterilize. The expiration date is marked on the outside of the package. The product should be stored in a cool, dry location. The 

instructions for use is recyclable. Dispose of used product and packaging following standard solid biohazard waste procedures. 
To remove the catheter from its tray, remove the handle retainer from the tray then lift the catheter up and out of the tray. See Figure 1. 

 

Figure 1.   

Warranty 

Abbott Medical warrants that its products shall be free from defects in materials and workmanship under normal use. This warranty does not exceed the 

"Expiration" date stated on any product labeling. The authorized uses and approved methods of use of each of our products is set forth in the related 

"Instructions for Use" that accompany each product. Abbott Medical disclaims any responsibility and liability for the use of its products in a manner that has 

not been authorized or approved. Abbott Medical’s liability under this warranty is limited to replacing its products. The foregoing warranty excludes and is in 

lieu of all other warranties whether expressed or implied including, but not limited to, any implied warranties of merchantability or fitness for a particular 

purpose. Abbott Medical disclaims any liability for any incidental or consequential loss, damage, or expense directly or indirectly arising from the use of this 

product, other than as expressly provided by specific law. Abbott Medical neither assumes nor authorizes any other person to assume for it any other or 

additional liability for loss, damage, or expense in connection with this product. For more details please review complete Abbott Medical warranty policy 

available from Abbott Medical or on the back of an Abbott Medical invoice. 

Symbols 

NOTE: The symbols section contains all the symbols that may be used on product labels. Product is labeled as required. 
Symbol 

Description 

 

Caution, consult accompanying documents 

 

Quantity 

 

Keep dry 

 

Keep away from sunlight 

 

Catalog number 

Summary of Contents for Advisor FL Circular Mapping Catheter, Sensor Enabled

Page 1: ...Advisor FL Circular Mapping Catheter Sensor Enabled Instructions for Use D AVSE D10 F20 D AVSE DF10 F20 D AVSE D10 F15 D AVSE DF10 F15 ...

Page 2: ...ornia to cause cancer and birth defects or other reproductive harm For more information go to www P65Warnings ca gov Indicates a trademark of the Abbott group of companies Indicates a third party trademark which is property of its respective owner Pat http www abbott com patents 2019 Abbott All Rights Reserved ...

Page 3: ...ccident cardiac damage perforation pericardial effusion or tamponade Risks associated with electrical stimulation may include but are not limited to the induction of arrhythmias such as atrial fibrillation AF ventricular tachycardia VT requiring cardioversion and ventricular fibrillation VF Do not use force to advance or withdraw catheter when resistance is encountered Do not immerse the proximal ...

Page 4: ...ng of the product to minimize product exposure to the atmosphere and to provide for aseptic product transfer It is recommended that the product remain in the unopened package until time of use Contents are sterile if the package is unopened and undamaged Do not resterilize The expiration date is marked on the outside of the package The product should be stored in a cool dry location The instructio...

Page 5: ...r use Follow instructions for use on this website Electrodes Spacing Number of electrodes Do not use if package is damaged Use by Length Temperature limitations Humidity limitation Outer Diameter Non pyrogenic Diagnostic Catheter Circular Mapping Catheter Conformité Européenne European Conformity Affixed in accordance with European Council Directive 93 42 EEC NB 2797 and 2011 65 EU Hereby Abbott M...

Page 6: ...2 ...

Page 7: ......

Page 8: ...lymouth MN 55442 USA 1 855 478 5833 1 651 756 5833 Abbott Medical 5050 Nathan Lane North Plymouth MN 55442 USA St Jude Medical Costa Rica Ltda Edificio 44 Calle 0 Ave 2 Zona Franca Coyol El Coyol Alajuela Costa Rica 2019 04 ARTEN600064427 A ...

Reviews: