SM3000
Multipoint Videographic Recorder
6 Configuration
IM/SM3000–EN Rev. L
53
Note.
The Internal Security Switch is used to access the Configuration level when 'Configuration security type' is set to 'Internal
switch protected' – see Section 6.4.2, page 65. Do Not use the switch to access the Configuration level when 'Configuration
security type' is set to 'Password protected' (default setting) unless the Password has been forgotten. The switch overrides
Password protection, enabling free access to the Configuration level.
Fig. 6.2 Setting the Security Switch
Note.
The red switch is for factory use only. Ensure it remains in the position closest to the rear of the instrument.
Withdraw the instrument
from the case
Remove tamper-proof
seal (if fitted)
Switch off the power to
the instrument and unscrew
the jacking screw securing
the instrument to the case
Set the yellow Security switch to the
'Configuration Level Not Protected' position
Refit the instrument to the case
and secure with the jacking screw
Configuration Level Not Protected
Configuration Level Protected
1
2
3
5
6
7
Re-apply power and wait for
the Operator level to display
Red Switch
(see
Note
below)
4
Access the Configuration level as
shown in Fig. 6.1, page 52