![ABB PGC5000 Series Service Instruction Download Page 6](http://html1.mh-extra.com/html/abb/pgc5000-series/pgc5000-series_service-instruction_3356407006.webp)
6
| DBDID | 892J009MNAE
Figure 2-1: PGC5000 Electronics compartment open
3
2
1
Legend: PGC5000 Electronics compartment open
Item
Description
1
DBDID amplifier assembly
2
DBDID power supply board
3
Flange nut
2.1
Detector removal
WARNING – Bodily injury
. High voltage present. Make sure that power to the PGC5000 has
been disconnected before removing the detector.
To remove the detector from the PGC5000, perform the following steps:
Disable the isothermal oven temperature zone and allow the oven to cool to room temperature.
Power down the oven.
Disconnect the column, cell vent line and the reactor gas inlet from the cell body. See items 1-
Figure 2-2: Disconnect lines
1
2
3
Summary of Contents for PGC5000 Series
Page 14: ...14 DBDID 892J009MNAE Figure 5 3 Diagram flow PGC5000 air and carrier...
Page 15: ...892J009MNAE DBDID 15 Figure 5 4 PGC5000 oven electronics internal connections diagram sheet 1...
Page 16: ...16 DBDID 892J009MNAE Figure 5 5 PGC5000 oven electronics internal connections diagram sheet 2...