Signal
Rated output value
at nominal load
Force
Sensitivity
F
nom
Figure 6. Sensitivity
Accuracy and Accuracy Class
Accuracy class is defined as the maximum deviation, and is expressed as a percentage of the sen-
sitivity at nominal load. This includes linearity deviation, hysteresis and repeatability error.
Linearity Deviation
Linearity deviation is the maximum deviation from a straight line drawn between the output values
at zero load and nominal load. Linearity deviation is related to the sensitivity.
Signal
Force
F
nom
Figure 7. Linearity Deviation
Hysteresis
Hysteresis is the maximum difference in the output signal at the same load during a cycle from zero
load to nominal load and back to zero load, related to the sensitivity at nominal load. The hysteresis
of a Pressductor transducer is proportional to the load cycle.
Signal
Force
F
nom
Figure 8. Hysteresis
Pressductor PillowBlock Load Cells, Vertical Measuring PFCL 201, User Manual
2 Description
12
3BSE023881R0101 en Rev H