106
| NG C820 0 AN D PGC1000 US ER MAN UA L | 210 1510MNAH
1.
Collect data from the device.
2.
Using the Lithium Battery Status instructions in section
, verify the battery status is OK
before proceeding.
NOTICE – Loss of data.
If the lithium battery is not sufficiently charged, a warm start will
become a cold start.
3.
Gain access to the rear termination panel on the analyzer by loosening the countersunk hex
socket locking set screw in the rear end cap using a 1/16” hex wrench and then unscrewing the
end cap.
4.
To warm start the device, depress the S1 reset button on the termination panel.
5.
Or, to remove the analyzer from service, disconnect the power connector J1 from the board.
6.
To place analyzer in service, return the power connection J1 to the termination panel.
10.9.2
Cold start
IMPORTANT NOTE:
A cold start will utilize factory calibration. Run manual peak find and
recalibrate the device after a cold start. See sections
DANGER – Explosion risk. Serious damage to health / risk to life
. Do not open or remove
covers unless the area is known to be non-hazardous.
To perform a cold start with the reset button:
1.
On the Analyzer Operation screen, click
Hold
under Next Mode. When the device completes the
current cycle and enters hold, continue to the next step.
2.
Gain access to the digital controller assembly by loosening the countersunk hex socket locking
set screw in the front end cap using a 1/16” hex wrench and then unscrewing the end cap.
3.
Gain access to rear termination panel on the analyzer by loosening the countersunk hex socket
locking set screw in the rear end cap using a 1/16” hex wrench and then unscrewing the end
cap.
4.
Unplug the lithium battery connector from the J5 receptacle on the digital controller board (see
).
5.
Push the Reset button located on the termination panel located in the rear of the enclosure.
6.
Initially, the Boot Loader screen will appear on the front display.
7.
When the Navigation screen appears, restore the lithium battery connection on the digital
controller board.
Figure 10-10: Digital controller complete assembly