Navigator 600 Silica
Single-stream
Appendix F – Updating the Software
84
IM/NAV6S/SS–EN Rev. E
Appendix F – Updating the Software
1. Unzip the upgrade software file and copy the contents of
the folder created to an SD card.
2. Ensure the analyzer is switched off.
3. Use a large flat-headed screwdriver to release the two
door catches
A
.
4. Remove the SD card
B
.
5. Set the switch
C
(above the SD card socket) to the lower
position (Software Update).
6. Insert the SD card, close the door and switch on the
analyzer. A start-up message is displayed for
approximately 5 seconds, followed by an 'ABB' splash
screen:
7. A 'Warning' screen appears next. This provides the option
to upgrade the existing system software, or exit without
upgrading:
8. To proceed, press the
key.
9. The 'Commissioning V2.0' screen appears with the
'Upgrade Firmware' button selected:
10. To upgrade the existing system software, press the
key to start the process. (The software is transferred to the
analyzer in sections.)
11. Press the
key as each section is installed (installation
may take a few moments).
12. Open the door and move the switch to the up position.
13. Close the door.
14. Turn power off and then on to restart the analyzer.
Fig. F.1 SD Card Removal
Note.
Use only genuine SanDisk SD cards with a
capacity of 2 Gb or less.
A
B
C
Normal
Operation (Up)
Software
Update
(Down)