Navigator 600 Phosphate
Multi-stream
5 Configuration
32
IM/NAV6P/MS–EN Rev. F
5.4.2 Streams
5.4.3 Diagnostics
Fields
Description
Stream 1 to 6
For each available stream a page opens with three fields:
Stream 1 (2 to 6) Tag – adds an identification for the current stream.
Range Low – sets the low range value (0 to 15 ppm).
Range High – sets the high range value (0 to 15 ppm).
Fields
Description
Reagent Bottle Capacity
The
Reagent Bottle Capacity
can be set between 0.1 and 10.0 liters. This should be changed only if
the initial volume of each reagent installed is not 2.5 liters.
This value is used to predict the level of reagent remaining in the bottle.
Reagent Limit
The early warning alarm point for the reagent solutions. Can be set between 1 and 30 days.
Set to a suitable interval to allow new reagents to be ordered.