ABB Navigator 550 Replacement Procedures Download Page 5

4 Replacement procedures

Navigator 550 | Low level dissolved oxygen wet-section | Spares replacement procedures |

INF12/113–EN Rev. C

5

4.5 Replacing the Modbus cable assembly

Modbus cable part numbers:

AW502 090 / 1.5 m (4.9 ft.)

AW502 091 / 5 m (16.4 ft.)

AW502 092 / 10 m (32.8 ft.)

AW502 093 / 20 m (65.6 ft.)

Referring to Section 4.2, page 3:

1. Remove the wet-section PCB cover.

Referring to Fig. 4.7:

2. Loosen cable gland

A

at the wet-section PCB housing

B

and disconnect the Modbus cable

D

from the terminal block

connections

C

marked black, red, screen, green and white.

3. Withdraw the Modbus cable

D

from the PCB housing and

main case assembly and discard.

4. Feed the replacement Modbus cable into the main case via

the guide channel

E

at the top, feed through the gland and

remake connections to the black, red, screen, green and

white PCB wet-section terminals and at the transmitter.

5. Refit the PCB cover by reversing the removal procedure.

Fig. 4.7 Replacing the Modbus cable assembly

B   R 

 SCR

G W

Summary of Contents for Navigator 550

Page 1: ...t made easy 1 Introduction These procedures must be carried out by a trained technician Tools required Pozidrive screwdriver Slot head terminal screwdriver Anti static strap 2 For more information Fur...

Page 2: ...olvents ELECTRICAL Isolate all high voltage supplies to the transmitter before performing replacement procedures The wet section is vulnerable to electrostatic damage Wear an anti static strap or dism...

Page 3: ...the new small PCB housing seal C between the PCB cover B and PCB housing F a small amount of silicone sealant or grease can be used to help retain the seal in place 11 Refit the PCB cover B using the...

Page 4: ...art numbers Drain valve assembly AW502 056 Drain valve cable assembly AW502 085 Referring to Fig 4 6 1 Depress the manual override button A on the drain valve B to drain the wet section 2 Disconnect t...

Page 5: ...cable gland A at the wet section PCB housing B and disconnect the Modbus cable D from the terminal block connections C marked black red screen green and white 3 Withdraw the Modbus cable D from the PC...

Page 6: ...Notes 6 INF12 113 EN Rev C Navigator 550 Low level dissolved oxygen wet section Spares replacement procedures Notes...

Page 7: ...Notes Navigator 550 Low level dissolved oxygen wet section Spares replacement procedures INF12 113 EN Rev C 7 Notes...

Page 8: ...n this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prio...

Reviews: