Navigator 550 | Hydrazine and sodium wet-sections | Flowmeter upgrade procedures (single-stream and multi-stream) |
INF13/141–EN Rev. B
5
Page 1: ...single stream and multi stream Measurement made easy 1 Introduction These procedures must be carried out by a trained technician Tools required Adjustable spanner 2 For more information Further inform...
Page 2: ...clean water Perform general cleaning of the wet section using a damp cloth only mild detergent can be used as a cleaning aid Do not use Acetone or any organic solvents ELECTRICAL Isolate all high vol...
Page 3: ...row on the flowmeter body is facing into the wet section 11 Remove and retain the 4 PCB cover fixing screws and washers Q and remove PCB cover R 12 Remove the upper cable gland blanking plug from the...
Page 4: ...and sodium multi stream wet sections sodium multi stream wet section shown R U Stream 3 Stream 2 sodium multi stream wet sections only Stream 1 Hydrazine and sodium single stream wet sections and use...
Page 5: ...Notes Navigator 550 Hydrazine and sodium wet sections Flowmeter upgrade procedures single stream and multi stream INF13 141 EN Rev B 5 Notes...
Page 6: ...rights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden with...