™ GLA232 Series Methane/Ethane Analyzer User Manual
ABB PROPRIETARY INFORMATION
37
Manually Adjust the Laser Offset
1. Click the
Setup
button on the
User Interface Control Bar
2. Select the
Laser Adjust
tab at the top of the screen. (Figure 22)
3. Check the
Disable Laser Frequency Lock
box to allow manual control of the laser.
4. Adjust the Laser A voltage using the
Laser A Frequency Offset
arrow buttons to shift the
peaks until they are centered on their respective target lines.
o
Up Arrow: Peaks adjust to the right
o
Down Arrow: Peaks adjust to the left
5. Adjust the Laser B voltage (bottom plot) using the
Laser B Frequency Offset
arrow
buttons to shift the peaks until they are centered on their respective target lines.
o
Up Arrow: Peaks adjust to the right
o
Down Arrow: Peaks adjust to the left
6. Uncheck the
Disable Laser Frequency Lock
box. The software resumes automatic
tracking and control of the laser wavelength.
7. Click
Close
to exit the menu and return to the
Main Panel
.
Serial Logger Tab
Use the
Serial Logger
tab (Figure 23) to configure the aggregation of data from additional serial
devices (for example, GPS and anemometer). This feature is configured as part of standard
MobileGuard™ sales and installations. Contact ABB for further support with this feature.
Figure 23: Serial Logger Tab