Parameterization
D184B097U02
FV4000 / FS4000
87
8.7.5 Standard
factor
Under constant operating conditions (constant pressure and temperature), the standard factor
can be entered here.
The standard factor is defined as the ratio between standard flow and operating flow:
Standard factor =
)
T
273
(
273
bar
013
,
1
)
p
bar
013
,
1
(
Qv
Qn
+
×
+
=
Since the mass flow is constant, the following formula applies:
N
V
Qv
Qn
ρ
ρ
=
Qn Standard flow
Qv Operating
flow
P
Operating pressure [bar ü]
T Temperature
[°C]
ρ
V Operating
density
ρ
N Standard
density
8.7.6 Operating
density
Please refer to the "AP-Calc" product selection and dimensioning program, which ABB provides
free of charge.