Endura AZ series integral and remote transmitter
Combustion oxygen monitor
8 Troubleshooting
88
IM/AZ20E–EN Rev. F
8.3 Performance Log
8.3.1 Log Entries
Information displayed in the
Performance Log
is derived from the values obtained during calibration
routines.
When the
Performance Log
is selected, an icon (
1
in Fig. 8.2) at the left side of the display indicates a
successful or failed calibration or accuracy check; or if a new probe / cell has been fitted. Icon types are
shown in Table 8.6 below:
Fig. 8.2 Typical Performance Log and Performance Code Structure
Item
Icon
1
Calibration check
successful
Calibration check
failed
Accuracy check
successful
Accuracy check
failed
New probe / cell
Table 8.6 Performance Log Icons
Item
Description
Item
Description
2
hh:mm:ss
YYYY–MM–DD
Time and date of entry
5
RR
Recovery rate
3
Off
Calibration offset
6
Dev
Deviation
4
Fct
Calibration factor
7
t
Time duration
Table 8.7 Performance Log Codes
hh:mm:ss
YYYY–MM–DD
Off x.xx
mV
Fct
x.xxx
RR xx.xx
%O2/s
Dev
xx.xx %O2
t
xxh xxm xxs
Back