![ABB EL3000 Series Manual Download Page 106](http://html1.mh-extra.com/html/abb/el3000-series/el3000-series_manual_3702924106.webp)
106
EL3000
CONTINUOUS GAS ANALYZERS | CI/EL3000-EN REV. B
Gas analyzer start-up
NOTE
The notes in the sections "ZO23: Gas analyzer start-up" (see page 107) and
"Fidas24: Gas analyzer start-up" (see page 108) must also be considered for
the ZO23 and Fidas24 analyzers.
Starting up the gas analyzer
1
Switch on the power supply of the gas analyzer.
2
The name of the gas analyzer and the number of the software version
are shown in the display while booting.
3
After the start-up phase has ended, the display switches over to the
measured value display.
Example:
4
Check the configuration of the gas analyzer and alter if necessary.
5
After the warm-up phase has ended, the gas analyzer is ready to carry
out measurements.
Duration of the warm-up phase:
Uras26: approx. ½ hour without thermostat, approx. 2 hours with ther-
mostat
Limas23: approx. 2 hours
Magnos206: approx. 1 hour
Magnos28: approx. 1 hour
Magnos27: approx. 2 to 4 hours
Caldos27: approx. ½ hour
Fidas24: approx. 2 hours
6
Check the calibration of the gas analyzer.
The gas analyzer is calibrated ex works. However, transport stresses and
the pressure and temperature conditions at the installation site may in-
fluence the calibration.
7
Turn on the sample gas supply.