GAA610-M
ADVANCED EMISSION GAS MONITORING SYSTEM | OI/GAA610-M-EN REV. B
… 12 Diagnosis / Troubleshooting
… Possible status messages
No.
Text
S A W F Reaction/Remark
Calibration (continuation)
508
Unknown error number. Check software versions.
Message during the automatic calibration. Check analyzer module and system
controller software versions.
509
Automatic calibration started.
For information
510
Automatic calibration ended.
For information
511
Automatic calibration interrupted externally.
For information
512
Automatic calibration in progress.
x For information; not logged
513
System bus communication faulty during
automatic calibration.
514
External calibration started.
For information
515
External calibration ended.
For information
516
External calibration in progress
x For information; not logged
517
Device being serviced.
x For information, e.g. during manual calibration, not logged
518
The calibration could not be performed, because
the measured value is not stable.
519
Preamplifier overflow error: Calibration could not
be performed because of preamplifier override.
520
Initial zero calibration started.
For information
521
Initial zero calibration ended.
For information
522
Initial zero calibration interrupted.
For information
523
Initial zero calibration incomplete. System bus
communication fault during calibration.
For information
524
Initial zero calibration started.
x For information; not logged
525
Linearization impossible: Linearization did not
produce a valid result. Measurement
See message text
526
Linearization impossible: Linearization could not
be performed, i.e. the characteristic is linear.
See message text
527
Initial calibration for component:
For information
528
Autocalibration not started, i.e. manual
calibration was running.
For information
529
Calibration was stopped because no raw
measured values were recorded.
x
x
530
Calibration stopped because the pressure switch
did not detect any calibration gas.
x
x
531
Automatic validation started.
For information
532
Automatic validation ended.
For information
533
Automatic validation externally interrupted.
For information
534
Automatic validation in progress.
x For information; not logged
535
Automatic validation successful for:
536
Automatic validation out of limits for:
537
Automatic validation out of limits for:
x