C1300
Advanced circular chart recorder
3 Configuration
IM/C1300 Rev. J
41
Totalizer Run Source
If required, select a digital signal to start the totalizer.
Refer to Appendix A on page 77 for a description of sources.
Totalizer Log Enable Source
If required, select the frequency at which the addition of the current totalizer values to the totalizer
log is triggered:
None
– current totalizer values are not added to the totalizer log
Prior to reset –
current totalizer values are added as the totalizer resets
12 hours
– current totalizer values are added every 12 hours
24 hours
– current totalizer values are added every 24 hours
Return to top of page.
Totalizer Log Start Time
Enter the time at which logging is to start.
Return to top of page.
Totalizer Secure Total
Secure Total Reset
Select
YES
to reset the secure total.
The secure total is reset to either 000000000 (
Count Direction
set to
Up
) or 999999999 (
Count
Direction
set to
Down
).
Return to top of page.
T1 Secure Reset
NO
Select Totalizer
T1 Secure Total
Select Totalizer
set to
T1 Secure Total
Tot 1 Run Source
None
Tot 1 Log Enable
None
Select Totalizer
T1 Configuration
Secure Total 1
000000000
gal
Tot 1 Log Start
00:00
None
Prior to reset
12 hours
24 hours
Select Totalizer
T1 Configuration