Aztec 600 aluminium, ammonia, color, iron, manganese, phosphate
Single- and multi-stream colorimetric analyzers
Appendix A – Reagents
78
IM/AZT6CR–EN Rev. J
A.3 Consumption of Sample, Standards and Dilution Water
The usage of sample and dilution water depends on the range
the analyzer is measuring and the number of sample rinses
selected.
Refer to Tables A.3 and A.4 to determine the sample and
dilution water usage per measurement. For High Level
Manganese, only 4 to 6 rinses should be allowed, for all other
parameters, only 0 to 4 rinses should be allowed.
The default cell rinse is 4 for high level manganese and 2 for all
others.
Refer to Table A.5 to determine the amount of de-ionised water
and standard used per calibration.
A.4 Cleaning Solutions
If a chemical cleaning cycle is run on the analyzer, the routine
uses approximately 8 ml of cleaning solution per cleaning cycle.
No. of Cell Rinses
Dilution
Ratio
0
1
2
3
1:0
7.5 ml
Sample
0 ml DI Water
17 ml
Sample
0 ml DI Water
26.5 ml
Sample
0 ml DI Water
36 ml
Sample
0 ml DI Water
1:1
3.75 ml
Sample
3.75 ml DI
Water
8.5 ml
Sample
8.5 ml
DI Water
13.25 ml
Sample
13.25 ml DI
Water
18 ml
Sample
18 ml
DI Water
1:2
2.5 ml
Sample
5.0 ml
DI Water
5.7 ml
Sample
11.3 ml DI
Water
8.85 ml
Sample
17.65 ml DI
Water
12 ml
Sample
24 ml
DI Water
1:3
1.90 ml
Sample
5.60 ml DI
Water
4.25 ml
Sample
12.75 ml DI
Water
6.6 ml
Sample
19.9 ml DI
Water
9 ml Sample
27 ml
DI Water
1:4
1.5 ml
Sample
6.0 ml DI
Water
3.4 ml
Sample
13.6 ml DI
Water
5.3 ml
Sample
21.2 ml DI
Water
7.2 ml
Sample
28.8 ml DI
Water
Table A.3 Sample and DI Usage per Measurement – 0 to 3
Rinses
No. of Cell Rinses
Dilution
Ratio
4
5
6
1:0
45.5 ml Sample
0 ml DI Water
55 ml Sample
0 ml DI Water
64.5 ml Sample
0 ml DI Water
1:1
22.8 ml Sample
22.8 ml DI Water
27.5 ml Sample
27.5 ml DI Water
32.3 ml Sample
32.3 ml DI Water
1:2
15.2 ml Sample
30.3 ml DI Water
18.3 ml Sample
36.7 ml DI Water
21.5 ml Sample
43.0 ml DI Water
1:3
11.4 ml Sample
34.1 ml DI Water
13.8 ml Sample
41.3 ml DI Water
16.1 ml Sample
48.4 ml DI Water
1:4
9.1 ml Sample
36.4 ml DI Water
11.0 ml Sample
44.0 ml DI Water
12.9 ml Sample
51.6 ml DI Water
Table A.4 Sample and DI Usage per Measurement – 4 to 6
Rinses
Number of Cell
Rinses
DI Water Usage
Calibration
Standard Usage
0
7.5 ml
7.5 ml
1
17 ml
17 ml
2
26.5 ml
26.5 ml
3
36 ml
36 ml
4
45.5 ml
45.5 ml
5
55.0 ml
55.0 ml
6
64.5 ml
64.5 ml
Table A.5 Calibration Solution Usage