12 Diagnostic Information and Icons
58
OI/AXM630–EN Rev. I
12 Diagnostic Information and Icons
12.1 Analyzer Diagnostic Information
The analyzer is programmed to display diagnostic messages to
provide information on servicing requirements and any other
conditions that develop during operation.
All diagnostic messages displayed on the analyzer are added to
the analyzer's audit log.
Table 12.1 shows icon types, diagnostic messages and
ON / OFF status for the fixed relays, all of which operate in the
fail-safe condition.
Note.
The diagnostic icons in the following table conform to
NAMUR 107.
Diagnostic Icon
NAMUR Status
Failure
Check function
Out of specification
Maintenance required
Icon
Diagnostic message
Stop
Relay
Attention
Relay
Calibration
Relay
Failure
Relay
Monitor in Service
The analyzer is measuring correctly.
ON
ON
ON
ON
20 Character User-defined Alarm Tag
User-defined alarm message.
—
—
—
—
Sample Flow Failure
The analyzer cannot detect any sample flow.
If there is sufficient flow to the analyzer, check that:
1.
the float is located in the side-sample pot and is unrestricted.
2.
the end cap of the side-sample pot is positioned to allow the
magnetic float to be within the area of operation of the reed switch.
3.
the float switch lead is connected correctly to the interconnection
board.
4.
the magnetic reed switch is operating correctly.
ON
OFF
ON
ON
Monitor Stopped
Analyzer operation has been stopped by the user.
OFF
ON
ON
ON
Monitor Off
Analyzer operation has not been started.
ON
ON
ON
ON
Monitor Off
A failure has caused the analyzer to stop operation and switch off all
services.
ON
ON
ON
OFF
Priming
The analyzer is priming the tubing. The priming routine draws in each
reagent, sample and standard in turn, filling the tubing and then pumping
to waste.
An automatic calibration starts when the priming routine is completed.
OFF
ON
ON
ON
Table 12.1 Diagnostic Information, Relay Operations and Inhibits (Sheet 1 of 4)