![ABB Aztec 600 Operating Instruction Download Page 51](http://html1.mh-extra.com/html/abb/aztec-600/aztec-600_operating-instruction_3446005051.webp)
Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers
7 Logging
OI/AXM630–EN Rev. I
49
7.5.3 Text Format Log Files (Audit and Alarm Log)
Alarm Event logs for each process group and the Audit log are
archived into individual files.
The filenames are formatted as shown in Table 7.6.
If one of the archive log files becomes full (>64000 entries) a new
file is created with an extension incremented by 1, for example:
a01, e01.
New text format log data files are also created when the daylight
saving period starts.
7.5.4 Daylight Saving
Files containing data generated during the daylight saving period
have '~DS' appended to the filename.
Start of daylight saving period
A daily file is started at 00:00:00 on 30
th
March 2012 filename:
30Mar12AW633.D00
Summertime starts at 2:00am on 30
th
March 2012 and the clock
changes automatically to 3:00am.
The existing file is closed and a new file is created filename:
30Mar12AW633~DS.D00
The file '30Mar12AW633.D00' contains data generated from
00:00:00 to 01:59:59.
The file '30Mar12AW633~DS.D00' contains data generated
from 03:00:00.
End of daylight saving period
A daily file is started at 00:00:00 on 26
th
October 2012 filename:
26Oct12AW633~DS.D00
Summertime ends at 3:00am on 26
th
October 2012 and the
clock changes automatically to 2:00am.
The existing file is closed and a new file is created filename:
26Oct12AW633.D00
The file '26Oct12AW633~DS.D00' contains data generated
from 00:00:00 to 02:59:59.
The file '26Oct12AW633.D00' contains data generated from
02:00:00.
7.5.5 Text Format Data Verification and Integrity
When text format data is saved to the archive media it is
checked automatically to verify that the data stored on the
media matches exactly what is stored in the internal memory.
7.6 Binary Format Data Files
Binary format archived data is stored in a secure binary encoded
format. A separate file is created for each recording channel.
The log data is stored in an encrypted text format.
The files can be read on a PC using ABB's DataManager Pro
data analysis software package.
7.6.1 Binary Format Data Filenames
When the 'Archive file format' parameter is set to 'Binary
format', the 'New File Interval' parameters (see Section 6.6.4,
page 40) are disabled and binary format filenames are created
with the content – see Table 7.7.
Filename extensions are assigned according to the data type
archived – see Table 7.8.
Log File
Filename
Alarm Event
<hour min> <day, mm, yy>
<process group tag>.e00
Audit Log
<hour min> <day, mm, yy>
<instrument tag>.a00
Table 7.6 Text Format Log File Format
Data Type
Filename Content
Stream Data
<Start Time HHMMSS>
<Start Date DDMMMYY>
Ch<Group><Channel><monitor tag>
e.g.: 14322719Mar12Ch1_2Final Water3
Alarm Event
Log Data
<Start Time HH_MM>
<Start Date DDMMMYY>
<Process Group Tag>
e.g.: 14_3219Mar12Final Water5
Audit log
Data
<Start Time HH_MM>
<Start Date DDMMMYY>
<Instrument Tag>
e.g.: 14_3219Mar12Final Water3
Table 7.7 Binary Format Data Filenames
Data
Binary
Format File
Extension
Stream data
*.B**
Alarm Event Log data files
– the historical record of the alarm events
related to the stream plus the history of any
operator messages
*.EE*
Audit Log data files
– the historical entries from the audit log
*.AE*