ABB AZ200 737 Instructions Manual Download Page 4

4

IN/ANAINST/032–EN Rev. B

| Endura AZ20 series probe | Combustion oxygen monitor

5 Removing the diffuser flame arrestor 
and cell

Referring to Fig. 5.1:
1.

At the probe cell end, loosen 6 M4 x 50 bolts 

A

 using an 

M4 spanner and 3 mm A/F hexagon wrench. Remove and 
discard the 6 M4 x 50 bolts 

A

 and associated nuts.

If the bolts are seized, use a small hacksaw to cut the 
through the bolts at the recess between the diffuser flame 
arrestor 

C

 and probe end plate 

B

.

2.

Withdraw diffuser flame arrestor 

C

 while carefully 

supporting cell 

D

. Discard diffuser flame arrestor 

C

.

If necessary, use a solvent, for example lighter fluid 

(petroleum ether) or alcohol (surgical spirits) to lubricate the 

joint between the diffuser flame arrestor and probe end.

3.

Remove test gas injection pipe 

E

 from the cell housing. 

Retain for re-assembly. 

4.

Carefully attempt to withdraw cell 

F

 from the cell housing 

using minimal force. Retain for re-assembly.
— If the cell can be withdrawn freely, proceed to step 6.
— If the cell does not move (is stuck in the probe body), 

use a spanner on the 2 flats to rock it sideways gently 

(no more than 10 to 15 °)

 until loose. If this action frees 

the cell proceed to step 6.

— If the cell cannot be withdrawn freely and / or is welded 

to the thermocouple / electrode contact assembly, 
proceed to step 5.

5.

If the cell is welded to the helical contact at the end of the 
thermocouple / electrode assembly, carefully withdraw the 
cell and thermocouple / electrode contact assembly until 
the cell tip 

G

 is visible. 

If the assembly will not withdraw sufficiently to access 
the joint between the cell and helical contact, disconnect the 
red (Cell +), green (TC+) and white (TC–) wires (at the probe 
head end) from the inner terminal block connections and 
straighten the wires – refer to probe Maintenance guide 

IM/AZ20M-EN

.

Use a sharp blade to separate the wire nest from the cell tip 

and separate the 2 items. 

6.

Remove and discard 'C-ring 

H

If the C-ring is stuck, insert a small screwdriver blade in the 
hollow of the C-ring 

I

 and gently prise it in several places 

until free.

7.

Proceed to Section 6, page 6 to fit the large surface area 
filter.

WARNING – Bodily injury

High temperature

When in operation, the cell end of the probe reaches 
temperatures up to 100 °C (212 °F). Allow the cell to 
reach ambient temperature before starting maintenance 
procedures.

IMPORTANT NOTE

Check all items for damage as they are removed. Do not 
re-use the nuts or bolts, always replace damaged items 
with new parts.

CAUTION – Potential damage to equipment

Do not use oil or release agents – these will damage 
the components.

CAUTION – Potential damage to equipment

Take extreme care not to damage either item when 
separating the cell tip from the wire nest on the 
electrode assembly.

CAUTION – Potential damage to equipment

Do not touch the sealing face or ID of the probe 
recess with the screwdriver blade. The surface finish 
must be undamaged to maintain the C-ring’s sealing 
properties.

Summary of Contents for AZ200 737

Page 1: ...e AZ20 probe These procedures must be carried out by suitably trained personnel and in accordance with the information given 2 Tools Items required M4 open ended spanner 3 mm A F hexagon wrench Solvent lighter fluid petroleum ether or alcohol surgical spirits Non metallic pan scouring pad Scalpel Small sharp blade Small flat bladed screwdriver Transmitter Instruction manual IM AZ20E EN Probe Instr...

Page 2: ...sor in a safe manner and in accordance with relevant local regulations Appropriate PPE including mask and goggles must be worn when preparing the process for these procedures WARNING Bodily injury Senor integral transmitter and transmitter weights are detailed in probe Instruction manual IM AZ20P EN Suitable lifting equipment must be available when installing removing the AZ20 components at the pr...

Page 3: ... the air supply and test gas connections at the sensor 2 Remove the sensor from the process in the reverse order of installation Ensure the sensor cable is free of cable ties restrictions before taking the sensor away from the process DANGER Serious damage to health risk to life Do not proceed until the transmitter is isolated from mains power supplies DANGER Serious damage to health risk to life ...

Page 4: ...rmocouple electrode assembly carefully withdraw the cell and thermocouple electrode contact assembly until the cell tip G is visible If the assembly will not withdraw sufficiently to access the joint between the cell and helical contact disconnect the red Cell green TC and white TC wires at the probe head end from the inner terminal block connections and straighten the wires refer to probe Mainten...

Page 5: ...Endura AZ20 series probe Combustion oxygen monitor IN ANAINST 032 EN Rev B 5 Fig 5 1 Removing the diffuser flame arrestor and cell 1 0 t o 1 5 m a x ...

Page 6: ...cate filter gaskets N ceramic filter O and clamp plate P 11 Pass 2 M4 x 60 screws Q through clamp plate P and fit 2 nuts R into the recess in ceramic filter adaptor K Tighten bolts evenly to a torque of 4 cNm 5 66 ozf in using a 3 mm A F hexagon wrench 12 Re install the probe as detailed in probe Instruction manual IM AZ20P EN 13 Make gas supplies to the probe and electrical supplies to the transm...

Page 7: ...Endura AZ20 series probe Combustion oxygen monitor IN ANAINST 032 EN Rev B 7 Fig 6 1 Refitting the cell and fitting the large surface area filter ...

Page 8: ...chnical changes or modify the contents of this document without prior notice With regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document We reserve all rights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to thi...

Reviews: