AW T 210 |
2- W I R E CO N D U C T I V I T Y, P H/O R P, P I O N T R A N SM I T T ER | O I/AW T 2 10 - EN R E V. C
33
2-electrode conductivity
Used to calibrate the sensor.
Note.
Calibrate
menus are sensor-specific – the menus in this section are applicable to
transmitters connected to 2-electrode conductivity sensor types.
Access to the
Calibrate
menu is permitted via the
Calibrate
and
Advanced
levels or directly from
an
Operator
page using the
Cal
button.
Menu
Comment
Default
Conductivity Cal
Enabled only if
Measurement Type = Conductivity.
Concentration Cal
Enabled only if
Measurement Type = Concentration.
Temperature Cal
Hold Output
HART versions only. Not available if HART in
Multidrop
or
Fixed Output
mode.
Disabled/Enabled.
When
Enabled:
output is held automatically at the beginning of a PV calibration and released automatically at the end.
When
Disabled:
output remains active throughout any PV calibration.
Edit Cal
PV Slope
PV Offset
Temp Slope
Temp Offset
Restore Defaults
4-electrode conductivity
Used to calibrate the sensor.
Note.
Calibrate
menus are sensor-specific – the menus in this section are applicable to
transmitters connected to 4-electrode conductivity sensor types.
Access to the
Calibrate
menu is permitted via the
Calibrate
and
Advanced
levels or directly from
an
Operator
page using the
Cal
button.
Menu
Comment
Default
Conductivity Cal
Enabled only if
Measurement Type = Conductivity.
Concentration Cal
Enabled only if
Measurement Type = Concentration.
Temperature Cal
Hold Output
HART versions only. Not available if HART in
Multidrop
or
Fixed Output
mode.
Disabled/Enabled.
When
Enabled:
output is held automatically at the beginning of a PV calibration and released automatically at the end.
When
Disabled:
output remains active throughout any PV calibration.
Edit Cal
PV Slope
PV Offset
Temp Slope
Temp Offset
Restore Defaults
Calibrate
Exit
Menu
Select
Calibrate
Exit
Menu
Select