Water Hardness Analyzer
AW101 Testomat
10 AW101 Testomat® Block Diagram
IM/AW101
Rev. C
19
10.1 Terminal Block Identification
No.
Terminal
Type
Function
Note
–
PE
IN
Mains – Protective earth (5x)
Earth/Ground
1 2
L N
IN
Mains, L = Live
Mains, N = Neutral
Mains input
115 V, 230 V or 24 V AC
3 to 5
6 to 8
n l
OUT
Neutral, switched (8x)
Live, switched (8x)
Mains voltage,
max. 4 A
9
10
11
LV1
OUT
Limit value output 1 – Normally closed
Limit value output 1 – Common
Limit value output 1 – Normally open
Volt-free relay output,
max. load 240 V AC, 4 A
12
13
14
LV2
OUT
Limit value output 2 – Normally closed
Limit value output 2 – Common
Limit value output 2 – Normally open
Volt-free relay output,
max. load 240 V AC, 4 A
15
16
17
Alarm
OUT
Fault message – Normally closed
Fault message – Common
Fault message – Normally open
Volt-free relay output,
max. load 240 V AC, 4 A
18
19
Stop 2
IN
External analysis stop Common earth for inputs
Only for volt-free normally
open/normally closed contact
20 21
IN 2
IN
Water meter input Common earth for inputs
Only for volt-free normally
open/normally closed contact
Note technical data of turbine
22 23
OUT + OUT –
OUT
0 or 4 to 20mA
Current output 22 + (0/4 to 20 mA)
23 –
24
+
OUT
+12 V for Hall-Sensor (turbine)
Note technical data of turbine. Max.
power input of sensor must not
exceed 20 mA
Fig. 10.2 Terminal Block Labels
n
n
n
l
l
l
PE PE
PE
L
N
Terminal block for mains connection and relay outputs
+
Stop
IN 2
OUT
+12
18 19 20 21 22 23 24
Terminal block for inputs
Stop, IN2 and output OUT
1
2
9 10 11
12 13 14
15 16 17
LV1
LV2
Alarm