- 22 -
Use PREV or NEXT key to scroll through the options and ENTER key to change or view the values.
The procedure to change the numeric value remains the one already explained for PASSWORD operation, i.e., the cursor starts
blinking on the most significant digit, then use the NEXT and PREV key to increase or decrease the value of the single digit (the
minus sign(-)automatically appears or disappears when the value increases above 9 or decreases below 0, as well as for the
decimal point(.). Use the ENTER key to move the cursor to the next digit, use the ESC key to move back to the previous digit.
An ENTER on the last digit will cause the value to be sent to the transmitter.
Refer to figures 7, 8, 9 and 10 for details.
Fig. 7 - CONF menu
Under ConF PERCENT option, the user can decide for linear
or SQR output. When SQR output is selected, the output is
linear from 0 to 20% (to 4% of input).
Refer to Fig. 6 - ConF METER for details on the procedures.
ConF XMTR - TRANSMITTER CONFIGURATION
Four are the operations under the ConF XMTR menu:
CONF, TRIM, REVIEW and PV.
By pressing ENTER on the ConF XMTR menu, the string
LOADING appears on the display, with the blinking star (*)
indicating communication activity, i.e. the CoMeter is reading
the transmitter information.
See below a list of the available operation under the selected option:
CONF menu
TRIM menu
REVIEW menu
PV menu
Change LRV
Reranging (RERANG.)
TAG 8
Primary variable (PRIMARY)
Change URV
Loop test (LOOPTST)
Final Assembly Nr. (XMTR N.)
Secondary variable (2ND)
Change DAMPING
Output trim (OUTTRIM)
Sensor Serial Nr. (SENS N.)
Tertiary variable (3RD)
Change UNITS
Zero adjustment (SNSZERO)
Up/Down scale (UP/DOWN)
Fourth variable (4TH)
Change OUTPUT
UNITS
LRV
URV
LRL (See Sensor Units)
URL (See Sensor Units)
DAMPING
OUTPUT
Then the CONF option appears.
Using PREV or NEXT key, the user can select CONF, TRIM,
REVIEW or PV option, and with the ENTER key he moves into
the menu.
When entering CONF and TRIM menu a message "LOOP
IN_MAN" appears to remind that a modification can change
the transmitter output, so for security the loop should be put in
Manual.
CONF
LOOP
IN_MAN.
ESC
ENTER
NEXT
PREV
ESC
ENTER
0.000
LRV
ESC
ENTER
00.000
01.000
01.000
01.000
01.000
LRV
ESC
ENTER
NEXT PREV
20.000
URV
ESC
ENTER
20.000
25.000
25.000
25.000
25.000
URV
ESC
ENTER
NEXT PREV
0.0000
DAMPING
ESC
ENTER
0.0000
0.2000
0.2000
0.2000
0.2000
DMP SEC
ESC
ENTER
NEXT PREV
ENTER
UNITS
ENTER
KPA
NEXT
PREV
ENTER
ESC
TORR
ATM
MPA
IN H2O
KG/CM2
OUTPUT
ENTER
LINEAR
ESC
SQR
NEXT PREV
NEXT
PREV
ESC
ESC
ESC
ESC
ENTER
ENTER
NEXT
PREV
ADDENDUM FOR COMETER - ANALOG LCD INDICATOR WITH HART PROGRAMMING CAPABILITY
NEXT PREV
TO MODIFY DIGIT,
DECIMAL POINT,
MINUS SIGN
NEXT PREV
TO MODIFY DIGIT,
DECIMAL POINT,
MINUS SIGN
NEXT PREV
TO MODIFY DIGIT,
DECIMAL POINT,
MINUS SIGN