SENSOR CAL.
– – – – –
0 0 0 0 0
Cal. User Code
Calibrating Zero
– – – – –
Fill Span Sol.
Calibrating Span
– – – – –
SECURITY CODE
0 0 0 0 0
0 0 0
1 2 0
.
0 0 0 0 0
Formazine Std.
Cal. Complete
– – – – –
DRY CAL. STD.
.
Zero Calibration
– – – – –
Calibration Fail
5
START UP AND OPERATION
…5.2.2
Calibration Page
Formazine Standard (wet) Calibration
– see also
Section 7.3.1
.
Model 7997 200 Series
Model 7997 300
Model 7997 400 Series
––––––
Advance to next parameter.
––––––
Calibration Access
Enter the required calibration code number, between 00000 and 19999. If an incorrect value
is entered, access to calibration is prevented and the Calibration Page is displayed.
–––––– Formazine (wet) Standard ––––––
Select
Formazine
Std.
by pressing
the
key.
Select
Formazine
Std.
by pressing
the
key.
Select
Formazine
Std.
by pressing
the
key.
––––––
Zero Calibration
––––––
Command to insert Dry
Calibration Standard Cell with
blank face towards the
receiver.
Command to fill chamber
with demineralized water.
Command to immerse the
sensor in de-mineralized
water.
After one of these actions, press
. The display changes to
Calibrating Zero
.
––––––
Calibrating Zero
––––––
Displayed for approximately
one minute, after which the
sequence changes to
Fill Span Sol.
Displayed for approximately
one minute, after which the
sequence changes to
Fill Span Sol.
Displayed for approximately
one minute, after which the
sequence changes to
Immerse
Fill Span Sol.
––– Span using Formazine Standard. –––
Use the
or
keys to set the value of the span solution being used.
Remove Dry Calibration
Standard and fill the inner
chamber with the Span
Solution.
Drain and fill with the Span
Solution
Remove system from
demineralized water and
immerse in the span solution.
Press
. The display changes to
Calibrating Span
.
Press
immediately to advance to
Cal. Complete
, if span calibration is not required.
––––––
Calibrating Span
––––––
Displayed for approximately a minute, after which the sequence changes to
Cal. Complet
e
(Calibration completed).
Cal. Complet
e
Calibration is now complete.
Cal. Fail
Cal Fail
indicates that the expected
step change between a zero and span
calibration was not produced by the
sensor. Check the Dry Standard,
calibration solutions, and repeat the
procedure.
Advance to
Access to Secure Parameters
– Section 6.1
.