Quality Control and Calibration
5-3
5.1.3
Entering Quality Control Values
Each lot of quality control material has assigned target values that must be entered into the HM5. This
can be done using any of three methods:
Enter values manually through the HM5’s touchscreen or a USB keyboard: see
Load USB control value files: see
“Loading Values from a USB Drive” on page 5-5
Load values through a USB 2D barcode reader: see
“Loading Control Values with a 2D
5.1.3.1
Entering Values Manually
Enter control values manually, for one level of control at a time, using
the HM5’s touchscreen or a USB keyboard.
1.
From the Home screen, select
Maintenance > Quality
Control
.
The Quality control screen includes buttons for QC levels
QC Low
,
QC Normal
, and
QC High
:
QC Normal
for normal controls (required)
QC Low
for low-level controls (optional)
QC High
for high-level controls (optional)
2.
Select a QC level.