Solares™ Smart Spa Bathing System Operating Manual
CD0036
- 8 -
1121
Input A.C. Power Port
Phenomenon
Basic EMC Standard
Immunity Test Levels
Electrical fast transients/ bursts
IEC 61000-4-4
± 2 kV
100 kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5
± 0.5 kV, ± 1kV, ± 2kV
Conducting disturbances
induced by RF fields
IEC 61000-4-6
3V
0.15 MHz - 80 MHz
6V in ISM bands between 0.15
MHz and 80 MHz
80 % AM at 1 kHz
Voltage dips
IEC 61000-4-11
0 % U
t
: 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°,
270°, and 315°
0 % U
t
: 1 cycle
and
70 % U
t
: 25/30 cycles
Single-phase at 0°
Voltage interruptions
IEC 61000-4-11
0 % U
t
: 250/300 cycle
Transportation and Storage Conditions
•
Ambient Temperature: -40°F to 158°F (-40°C to 70°C)
•
Relative Humidity Range: 10% to 100%
•
Atmospheric Pressure Range: 800 to 1060 hPa
Operation Conditions
•
Ambient Temperature: 75°F to 95°F (24°C to 35°C)
•
Relative Humidity Range: 10% to 100%
•
Atmospheric Pressure Range: 800 to 1060 hPa
Identification Label Location
The label describing the model #, Serial #, and electrical information is located behind the panel on the
right side of the tub. Remove the panel and kneel next to the right side of the
tub to review the label’s
information. (Label is on front of the upper frame.)
Intended Use
Apollo bathing systems are designed and sold for use in medical field markets including but not limited
to nursing homes, hospitals, and assisted living facilities. Apollo’s bathing systems are designed to be
operated by trained operators and are not intended to be operated by the individual taking the bath.
The device is intended to be operated with the resident to be bathed seated in the tub and the operator
at either side.