a&p instruments 2001 Series Instruction Manual Download Page 7

 

7

Please  note  that  the  standard  settings  will  be  correct  for  the  most  materials  with 
sample  sizes  of  approx.  10  g,  which  is  our  recommendation  as  a  good  sample 
weight. 
 

3

) “

PFl

” is the 

“PreFlash

” feature, which is useful for very wet samples to shorten the 

measuring  time.  This  parameter  is  set  to  a  certain  value  for  the  remaining  dry 
substance.  E.g.  a  material  has  moisture  of  90  %  and  dry  material  of  10  %,  the 
parameter  should be  set  to  30  %.  Then  the  radiator  will  start  to  dry  with  full energy 
(100  %)  and  when  the  value  of  30  %  dry  substance  is  reached,  it  will  dry  to  the 
automatic  stop  with  a  lower  energy  set  by  the  parameter  “Heat”.  Results  has  to  be 
set to “dM” in that case. Users in waste water plants often use this feature. 
 
 

5. Drying temperature 

 
The  drying  temperature  can  be  set  by  adjusting  the  menu  item  “Heat”  to  an  exact 
value  of  %  energy.  This  means  %  of  230  Volt  power.  Of  course  there  is  a  certain 
connection between the power and the temperature.  
 

Radiation power in % and sample temperature (250 W radiator) 

 

100 %   160 – 170

0

 C 

90 %     140 – 150

0

 C 

80 %     130 – 140

0

 C 

70 %     120 – 130

0

 C 

60 %     110 – 120

0

 C 

50 %     85 – 95

0

 C 

40 %     70 – 80

0

 C 

30 %     55 – 65

0

 C 

20 %     45 – 50

0

 C 

10 %     40 – 45

0

 C 

Please take into consideration, that the sample temperature is depending very much 
on the quality and mostly on the colour of the sample. A light sample will get much 
less hot than a dark sample. Therefore the temperatures are only approximate. 
 
The radiator’s power can be changed also during a  measurement. Press the „Tara“ 
button and the „Set“ button at the same time. Use the turn switch to set the power in 
%. The value is shown in the display. 
 
(For UX 2031 models only) 
Only the bright radiator is adjustable, the quartz radiator dries only with 100 % power. 
 
 

Summary of Contents for 2001 Series

Page 1: ...Instruction Manual Series 2001 instruments p a a p instruments e K Kluckhofer Weg 61 D 32657 Lemgo Tel 49 05232 9778 0 Fax 49 05232 9778 20 Email ap instruments t online de Internet www apinstruments...

Page 2: ...n a vibration free surface Parts of sample material shall not fall between the pan holder and the housing that will hinder the balance Never touch the hot radiator there is a risk of injury Pull out t...

Page 3: ...20 sec interval 1 min start interval That means the instrument will start the drying process wait 1 minute after the start start interval until the weight loss is registered by the processor and it wi...

Page 4: ...e instruments into the automatic mode 0 000 g press SET 2 seconds CON 12 3 or press change 12 3 blinks set time with button change confirm with SET 10 3 blinks 10 0 blinks press SET T I M E R is s e t...

Page 5: ...e Settings level DISPLAY shows SETTINGS can be changed by using the Change button MEANING PNr 0000 1000 9000 Standard is no sample no Sample no 0001 0002 IA 0 6 12 30 Sek 1 2 4 min Printing in interva...

Page 6: ...accuracy of measurement of approx 0 1 this parameter has to be set correctly E g a sample weight of approx 4 8 g needs a setting to 4 mg because 4 mg is 0 1 of 4 g Sample weight Recommended setting 4...

Page 7: ...drying temperature can be set by adjusting the menu item Heat to an exact value of energy This means of 230 Volt power Of course there is a certain connection between the power and the temperature Ra...

Page 8: ...weight loss even before the start of the measurement 7 4 We recommend a sample weight of at least 4 g The greater the sample weight the better reproducible your measurement will be 7 5 Try to find ou...

Page 9: ...contact with the housing The instrument shows other irregular reactions e g the radiator cannot be regulated or the parameters are lost Switch off and on the instrument Check if the correct type of i...

Page 10: ...xistent The data correspond to the V 24 agreements for serial asynchronous data transmission The structure of a data word follows from the following sketch Start Start LSB MSB Stop 0V 12V Bit 0 1 6 Bi...

Reviews: