English 11
Table 1 - EMISSION Limits -
Phenomenon
Compliance
Conducted and radiated RF EMISSION CISPR 11
Group 1 Class B
Harmonic distortion
IEC 61000-3-2
Class A
Voltage fluctuations and flicker
IEC 61000-3-3
Compliance
Table 2 - IMMUNITY TEST LEVELS : Enclosure Port -
Phenomenon
IMMUNITY TEST LEVELS
Electrostatic discharge
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated RF EM fields
IEC 61000-4-3
10 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
Proximity fields from RF wireless communications
equipment IEC 61000-4-3
See table 4
Rated power frequency magnetic fields
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
Table 3 - IMMUNITY TEST LEVELS : Input a.c. power Port -
Phenomenon
IMMUNITY TEST LEVELS
Electrical fast transients / bursts
IEC 61000-4-4
±2 kV
100 kHz repetition frequency
Surges Line-to-line IEC 61000-4-5
±0.5 kV, ±1 kV
Conducted disturbances induced by
RF fields IEC 61000-4-6
3 V 0.15 MHz - 80 MHz
6 V in ISM and amateur radio bands between 0.15
MHz and 80 MHz 80 % AM at 1 kHz
Voltage dips IEC 61000-4-11
0 % U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
0 % U
T
; 1 cycle And 70 % U
T
; 25/30 cycle
Single phase: at 0°
Voltage interruption IEC 61000-4-11 0% U
T
; 250/300 cycle
NOTE U
T
is the AC mains voltage prior to application of the test level.