background image

 

English   8 

 

6.  Using the Monitor 

Installing / Changing the Batteries 

1.  Remove the battery cover.   
2.  Insert a new set of batteries into the battery 

compartment as shown. Make sure the polarities 
(+) and (-) are correct. Use only LR6, AA or 
equivalent batteries.   

3.  Close the battery cover. 
 

CAUTION 

  Insert the batteries in the battery compartment. 

If not, the device will not work. 

 When 

 (LOW BATTERY mark) blinks in the LCD display, replace all 

batteries with new ones. Do not mix old and new batteries. It may shorten the 
battery life, or cause the device to malfunction. 

  Battery life varies with the ambient temperature and may be shorter at low 

temperatures. 

  Use the specified batteries only. The batteries provided with the device are for 

testing monitor performance and may have a limited life. 

  Remove the batteries if the device will not be in use for a prolonged period of 

time as the batteries may leak and cause a device malfunction. 

 
 

Connecting the Air Tube 

 

Insert the air tube into the cuff connector firmly. 

 
 

Selecting the Proper Cuff 

 

Using the correct cuff size is important for accurate readings. If the cuff is not 
the proper size, the reading may yield an incorrect blood pressure value.   

  The INDEX and RANGE marking on the cuff will indicate that you are using 

the proper cuff or not. (Refer to "

Attaching the Arm Cuff

" in next section) 

 

Arm Size 

Recommended Cuff Size 

Order No. 

33 cm to 45 cm 

Large cuff 

AX-UM-101-03 

23 cm to 33 cm 

Medium cuff 

AX-UM-101-01 

17 cm to 25 cm 

Small cuff 

AX-UM-101-02 

 

Arm size: The circumference at the biceps 

Summary of Contents for UM-101

Page 1: ...Instruction Manual Manuel d instructions Manual de instrucciones Manuale di Istruzioni Bedienungsanleitung WM PD4001301...

Page 2: ...uction Manual Contents 1 Dear Customers 2 2 Preliminary Remarks 3 3 Notes for Proper Use 4 4 Parts Identification 6 5 Symbols 7 6 Using the Monitor 8 7 Measurements 9 8 Useful Features 10 9 Troublesho...

Page 3: ...sphygmomanometer is designed to monitor and display the cuff pressure during cuff inflation and deflation while the healthcare provider determines the patient s blood pressure level by listening for...

Page 4: ...e mark of conformity accompanied by the reference number to the involved notified body This device is designed for adults only Definitions SYS Systolic Blood Pressure DIA Diastolic Blood Pressure PUL...

Page 5: ...her liquids If the device is immersed by accident it may require servicing DO NOT use the device before it has endured a full service Do not leave the device in a high temperature or high humidity env...

Page 6: ...ccessories before any subsequent uses Do not pull or kink the tubes Do not use any organic solvent for example antiseptic solution etc to clean the device Press the ON OFF button after a measurement Y...

Page 7: ...During measurement After exhausted air Indicates pressure value Indicates pulse rate Display Numerical display Rubber bulb Exhaust valve ON OFF button MARKER button Pressure bar display Cuff 1 5V Batt...

Page 8: ...are designed to provide special protection against electrical shocks Full Battery The battery power indicator during measurement Low Battery The battery is low when it blinks Replace all batteries wit...

Page 9: ...ies only The batteries provided with the device are for testing monitor performance and may have a limited life Remove the batteries if the device will not be in use for a prolonged period of time as...

Page 10: ...lashing the device is ready for measurement If air is remaining in the cuff when the ON OFF button is pressed the display will indicate an error code Switch off the device press the ON OFF button agai...

Page 11: ...An error message will be displayed if a measurement is taken with insufficient pulses or in a very noisy environment 7 Press the ON OFF button again to turn off the power Note Model UM 101 has an aut...

Page 12: ...culty in determining thier blood pressure The unit does not take a measurement Readings are too high or too low Remove the batteries Reinstall them properly and try the measurement again Note If the r...

Page 13: ...Pressure 3 mmHg Pulse 5 Pressure bar display Pressure 4 mmHg Power supply 2 x 1 5 V alkaline batteries LR6 or AA Upper arm circumference 23 33 cm using the medium cuff Classification Type BF EMC IEC...

Page 14: ...LTD Authorized Representative Established in the European Community Unit 24 26 Blacklands Way Abingdon Business Park Abingdon Oxon OX14 1DY United Kingdom Telephone 44 1235 550420 Fax 44 1235 550485...

Reviews: