Digital Blood Pressure Monitor 

 
 

Model UA-852

 

 

Instruction Manual 

 
 

Manuel d

instructions 

 
 
 

Manual de Instrucciones 

 
 
 

Manuale di Istruzioni 

 
 
 

Bedienungsanleitung

 
 
 
 

使用手册

 

1WMPD4001655B

 

Summary of Contents for UA-852

Page 1: ...Digital Blood Pressure Monitor Model UA 852 Instruction Manual Manuel d instructions Manual de Instrucciones Manuale di Istruzioni Bedienungsanleitung 1WMPD4001655B...

Page 2: ...o Take Proper Measurements 8 Measurement 8 After Measurement 8 Measurements 9 Normal Measurement 9 Measurement With The Desired Systolic Pressure 10 Notes For Proper Measurement 10 Recalling Memory Da...

Page 3: ...r use The device is for use indoors Precautions Precision components are used in the construction of this device Extremes in temperature humidity direct sunlight shock or dust should be avoided Clean...

Page 4: ...e illumination Air Connector Plug Air Socket MEMORY Button Part of Display Pressure Bar Indicator and WHO Classification Indicator MEMORY Systolic Pressure Pulse Rate Battery Indicator I H B Indicator...

Page 5: ...EMORY FullBattery The battery power indicator during measurement LowBattery The battery is low when it blinks Replace all batteries with new ones when the indicator blinks Unstable blood pressure due...

Page 6: ...EMORY button is pressed the memory data is displayed from the most recent For details on recalling data refer to Recalling Memory Data on page 11 The device can store the last ninety sets of data in m...

Page 7: ...OW BATTERY mark does not appear when the batteries are drained Battery life varies with the ambient temperature and may be shorter at low temperatures Use the specified batteries only The batteries pr...

Page 8: ...e to hour minute selection 5 Select the hour using the button Press the CLOCK button to set the current hour and move to minute selection 6 Select the minute using the button Press the CLOCK button to...

Page 9: ...on the table Relax for about five to ten minutes before measurement Place the center of the cuff at the same height as your heart Remain still and keep quiet during measurement Do not measure right af...

Page 10: ...op inflation at any time press the START button again 3 When inflation is complete deflation starts automatically and heart mark blinks indicating that the measurement is in progress Once the pulse is...

Page 11: ...d or depressed by emotional stress the measurement will reflect this stress as a higher or lower than normal blood pressure reading and the pulse reading will usually be faster than normal An individu...

Page 12: ...is displayed 3 Then each time the MEMORY button is pressed the previous data is displayed 4 After the last data is displayed the display returns to the clock display Clearing the memory data 5 Press...

Page 13: ...n if you see this indicator frequently Pressure Bar Indicator The indicator monitors the progress of pressure during measurement WHO Classification Indicator Each segment of the bar indicator correspo...

Page 14: ...regular physical checkups Maintain proper weight Why Measure Blood Pressure At Home Blood pressure measured at a clinic or doctor s office may cause apprehension and can produce an elevated reading 2...

Page 15: ...etely the mark does not appear Replace all batteries with new ones The cuff is not fastened properly Fasten the cuff correctly You moved your arm or body during the measurement Make sure you remain ve...

Page 16: ...ta Type UA 852 Measurement method Oscillometric measurement Measurement range Pressure 20 280 mmHg Pulse 40 180 beats minute Measurement accuracy Pressure 3 mmHg or 2 whichever is greater Pulse 5 Powe...

Page 17: ...age power supply network that supplies buildings used for domestic purposes Recommended separation distances between portable and mobile RF communications equipment and the A D unit The A D unit is in...

Page 18: ...F transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the...

Page 19: ...ential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage var...

Page 20: ...gdon Oxfordshire OX14 1DY United Kingdom Telephone 44 1235 550420 Fax 44 1235 550485 A D Australasia Pty Ltd 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 83...