5
Symbols
Symbols
Function / Meaning
Treatment
Turns on or off the device
Sets the clock / Does not store the
current data
Direction guide to install battery
Direct current
SN Serial
number
Date of manufacture
Type BF: Device, cuff and tubing are
designed to provide special protection
against electric shocks.
Appears when measurement is in
progress and flashes when the pulse
is detected.
Measurement is in progress --
remain as still as possible.
Previous measurement stored in
MEMORY
Appears when the battery voltage is
too low for the device to work properly.
Replace all batteries with new
ones.
Appears if the systolic and diastolic
measurements are within 10 mmHg of
each other.
Verify that the air hose is
properly connected and that the
exhaust rate is between 2 and 5
mmHg. Check for air leakage.
Appears if the pressure value is
unstable due to movement during the
measurement.
Try the measurement again.
Remain very still during the
measurement.
Appears if the pressure value did not
increase during pump up.
Verify that the air hose is
properly connected and that the
exhaust rate is between 2 and 5
mmHg. Check for air leakage.
System error
Reload batteries.
Appears if pulses are not detected
correctly.
Attach the cuff properly and try
the measurement again.
Attention symbol
See instruction for use
EC directive medical device label
WEEE label