611 PLUS Blood Pressure Monitor
Congratulations on purchasing a state-of-the-art A&D blood pressure monitor, one of
the most advanced monitors available today. Designed for ease of use and accuracy, this
device will facilitate your daily blood pressure regimen. We recommend that you read
through this manual carefully before using the device for the first time.
PRELIMINARY REMARKS
☞
This device conforms to the European Directive 93/42 EEC for Medical Products.
This is made evident by the
mark of conformity. (0123: The reference
number to the involved notified body)
☞
The device is designed for use on adults only. Do not use on newborns or
infants.
☞
This device is intended for home healthcare environment.
☞
This device is designed to measure blood pressure and pulse rate of people for
diagnosis.
PRECAUTIONS
Precision components are used in the construction of this device. Extremes in
temperature, humidity, direct sunlight, shock or dust should be avoided.
Clean the device and cuff with a dry, soft cloth or a cloth dampened with water and
a neutral detergent. Never use alcohol, benzene, thinner or other harsh chemicals to
clean the device or cuff.
Avoid tightly folding the cuff or storing the hose tightly twisted for long periods, as
such treatment may shorten the life of the components.
The device and cuff are not water resistant. Prevent rain, sweat and water from
soiling the device and cuff.
Measurements may be distorted if the device is used close to televisions, microwave
ovens, cellular telephones, X-ray or other devices with strong electrical fields.
When reusing the device, confirm that the device is clean.
Used equipment, parts and batteries are not treated as ordinary household waste,
and must be disposed of according to the applicable local regulations.
Do not modify the device. It may cause accidents or damage to the device.
To measure blood pressure, the arm must be squeezed by the cuff hard enough to
temporarily stop blood flow through the artery. This may cause pain, numbness or a
temporary red mark to the arm. This condition will appear especially when
measurement is repeated successively. Any pain, numbness, or red marks will
disappear with time.
People who have a severe circulatory deficit in the arm must consult a doctor before
using the device, to avoid medical problems.
Do not self-diagnose the measurement results and start treatment by yourself.
Always consult your doctor for evaluation of the results and treatment.
Do not apply cuff on an arm with other medical equipment attached. Neither
equipment may function properly.
Do not apply the cuff on an arm with an unhealed wound.
Do not apply the cuff on an arm receiving an intravenous drip or blood transfusion.
It may cause injury or accidents.
Do not use the device where flammable gases such as anesthetic gases are present. It
may cause an explosion.
Do not use the device in highly concentrated oxygen environments, such as a high-
pressure oxygen chamber or an oxygen tent. It may cause a fire or explosion.
Do not excessively bend or squeeze the air hose.
MONITOR COMPONENTS
DISPLAY
SYMBOL IDENTIFICATION
Symbols printed on the device case:
DISPLAY SYMBOLS
Symbols that appear on the display:
ABOUT BLOOD PRESSURE
What Is Blood Pressure?
Blood pressure is the force exerted by blood against the walls of the arteries.
Systolic pressure occurs when the heart contracts. Diastolic pressure occurs
when the heart expands. Blood pressure is measured in millimeters of
mercury(mmHg).
INSTALLING BATTERIES
1. Remove the battery cover.
2. Remove the used batteries from the battery
compartment in case of changing them.
3. Insert new batteries into the battery compartment
as shown, taking care that
the polarities (+) and (-) are correct.
4. Replace the battery cover.
Use only R6P, LR6 or AA batteries.
CAUTION
Insert the batteries as shown in the battery compartment. If installed incorrectly, the
device will not work.
When
(LOW BATTERY mark) blinks on the display and the device announces
that the battery needs to be replaced, replace all batteries with new ones. Do not mix
old and new batteries. It may shorten the battery life, or cause the device to
malfunction.
Replace the batteries two seconds or more after the device turns off.
If
(LOW BATTERY mark) does not appear then the batteries are drained.
The battery life varies with the ambient temperature and may be shorter at low
temperatures.
Remove the batteries if the device is not to be used for a long time.
The batteries may leak and cause a malfunction.
Index
Point: Cuff fit range indicator (printed on the cuff)
CUFF SYMBOL IDENTIFICATION
Symbols printed on the cuff:
APPLYING THE ARM CUFF
1. Wrap the cuff around the upper arm, about 1 to
2 cm above the inside of the elbow, as shown.
Place the cuff directly against the skin, as clothing
may cause a faint pulse and result in a
measurement error.
2. Constriction of the upper arm, caused by rolling
up a shirtsleeve, may prevent accurate readings.
3. Confirm that the index
points within the
proper fit range.
HOW TO TAKE ACCURATE MEASUREMENTS
For the most accurate blood pressure measurement:
Sit comfortably on a chair. Rest your arm on the table. Do not cross your legs. Keep
your feet on the floor and straighten your back.
Relax for several minutes before measurement.
Place the center of the cuff at the same level as your heart.
Remain still and keep quiet during measurement.
Do not measure immediately after physical exercise or a bath. Rest for twenty or
thirty minutes before taking the measurement.
Try to measure your blood pressure at the same time every day.
DURING MEASUREMENT
During measurement, it is normal for the cuff to feel very tight. (Do not be alarmed).
AFTER MEASUREMENT
After measurement, press the START button to turn the device off. After one
minute of non-operation, the device will turn off automatically.
Remove the cuff and record your data.
TAKING A MEASUREMENT
Model UA-611 is designed to detect the pulse
and to inflate the cuff to a systolic pressure level
automatically.
If your systolic pressure is expected to exceed 230
mmHg, read “Measurement With A Desired Systolic
Pressure” below.
Normal Measurement
1. Place the cuff on the arm (preferably the left arm).
Sit quietly during measurement.
2. Press the button. All of the display segments
are displayed. Zero is displayed blinking briefly.
Then the display changes, as indicated in the
figure at the right, as the measurement begins. The
cuff starts to inflate. It is normal for the cuff to feel
very tight. A pressure bar indicator is displayed, as
in the figure at the right, during inflation.
Note: If you wish to stop inflation at any time, press the
button again.
3. When inflation is complete, deflation starts
automatically and the
(heart mark) blinks,
indicating that the measurement is in progress.
Once the pulse is detected, the mark blinks with
each pulse beat.
Note: If an appropriate pressure is not obtained, the device
starts to inflate again automatically.
4. When the measurement is complete, the systolic
and diastolic pressure readings and pulse rate are
displayed. The cuff exhausts the remaining air and
deflates completely.
5. Press the button again to turn off the power.
Note: Model UA-611 is provided with an automatic power
shut-off function.
M
Measurement with a Desired Systolic Pressure
If your systolic pressure is expected to exceed 230
mmHg, use this procedure.
1. Place the cuff on the arm (preferably the left
arm).
2. Press the button.
3. During the zero blinks, press and hold the
button until a number about 30 to 40 mmHg
higher than your expected systolic pressure
appears.
4. When the desired number is reached, release
the button to start measurement. Then
continue to measure your blood pressure as
described in the previous section.
TIPS FOR TAKING READINGS
Sit down in a comfortable position. Place the arm to be used for the measurement on
a table or other support so that the center of the cuff will be at the same height as
your heart.
Relax before taking a measurement. If you are excited or depressed by emotional
stress, the measurement will reflect this stress as a higher (or lower) than normal
blood pressure reading and the pulse reading will usually be faster than normal.
An individual’s blood pressure varies constantly, depending on what you are doing
and what you have eaten. What you drink can have a very strong and rapid effect on
your blood pressure.
This device bases its measurements on the heartbeat. If you have a very weak or
irregular heartbeat, the device may have difficulty determining your blood pressure.
Should the device detect a condition that is abnormal, it will stop the measurement
and display an error symbol. Refer to the section titled Display Symbols for the
description of the symbols.
This device is intended for use by adults only. Consult with your physician before
using this device on a child. A child should not use this device unattended.
RECALLING THE MEMORY
The device automatically stores up to fifteen blood pressure and pulse measurements in
memory. Data stored in memory are assigned a data number in the order of the newest
to the oldest. The oldest data displays as “
“. The M symbol in the upper left
corner of the display indicates that you are viewing previous data stored in memory.
1. When nothing is displayed, press and hold
the button to recall the stored data.
2. Release the button when displaying the
average data.
3. The data number and stored data are
automatically displayed in order from the last
measurement.
4. The display will turn off automatically after
all data is displayed.
Note: If you press the button while recalling
data, the device turns off.
DELETING THE MEMORY
1. When turning off the device, press and hold
the button until the
“
“ is displayed.
2. Select “
“ to clear data.
3. Data is cleared when the
mark blinks.
4. The device turns off automatically.
TROUBLESHOOTING
Note: If the actions described above do not solve the problem, contact the dealer. Do not attempt to open
or repair this product, as any attempt to do so will make your warranty invalid.
M
AC Adapter port
UA-651
JNC7
JNC7
MAINTENANCE
Do not open the device. It uses delicate electrical components and an intricate air
unit that could be damaged. If you cannot fix the problem using the troubleshooting
instructions, contact the authorized dealer in your area or our customer service
department. A&D customer service will provide technical information, spare parts and
units to authorized dealers. It is generally recommended to have the monitor inspected
every 2 years to ensure proper functioning and accuracy.
SPECIFICATIONS
MODEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .UA-611
MEASUREMENT METHOD . . . . . . . . . . . . . .Oscillometric measurement
MEASUREMENT RANGE . . . . . . . . . . . . . . . .Pressure: 0 - 299 mmHg
Systolic pressure: 60 - 279 mmHg
Diastolic pressure: 40 - 200 mmHg
Pulse: 40 - 180 beats / minute
MEASUREMENT ACCURACY . . . . . . . . . .Pressure: ±3 mmHg
Pulse: ±5 %
POWER SUPPLY . . . . . . . . . . . . . . . . . . . . . . . .4 x 1.5V batteries (R6P, LR6 or AA)
NUMBER OF MEASUREMENTS . . . . . . . . . .Approx. 700 times LR6 (alkaline batteries)
Approx. 200 times R6P (manganese batteries)
With pressure value 180 mmHg, room
temperature 23 °C
CLASSIFICATION . . . . . . . . . . . . . . . . . . . . . . .Internally powered ME equipment (Supplied
by batteries). Continuous operation mode.
CLINICAL TEST . . . . . . . . . . . . . . . . . . . . . . . .According to ANSI / AAMI SP-10 1992
EMC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .IEC 60601-1-2: 2007
MEMORY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Last 15 measurements
OPERATING CONDITIONS . . . . . . . . . . . . . .50 °F to 104 °F / 15 %RH to 85 %RH
+10 °C to +40 °C / 15 %RH to 85 %RH
800 to 1060hPa
TRANSPORT/STORAGE CONDITIONS . . .-4 °F to 140 °F / 10 %RH to 95 %RH
-20 °C to +60 °C / 10 %RH to 95 %RH
DIMENSIONS. . . . . . . . . . . . . . . . . . . . . . . . . . .Approx. 3.8" (W) x 2.7” (H) x 5.1” (D)
Approx. 96 (W) x 68 (H) x 130 (D) mm
WEIGHT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Approx. 8.1 oz., excluding the batteries
Approx. 230 g, excluding the batteries
INGRESS PROTECTION . . . . . . . . . . . . . . . . .Device: IP21
APPLIED PART . . . . . . . . . . . . . . . . . . . . . . . . .Cuff Type BF
USEFUL LIFE . . . . . . . . . . . . . . . . . . . . . . . . . . .Device: 5 years (when used six times a day)
Cuff: 2 years (when used six times a day)
Note: Specifications are subject to change without prior notice.
EMC table information is listed on our website:
http://www.aandd.jp/products/manual/medical/emc_en.pdf
Accessories sold separately
Part Number
Cuff Size
Arm Size
UA-420
AccuFit
TM
Plus Cuff
8.6"-16.5" (22-42cm)
UA-290
Medium Cuff
9.0"-14.6" (23-37cm)
UA-291
Large Cuff
12.2"-17.7" (31-45cm)
Arm size: The circumference at the biceps.
WARRANTY
This blood pressure monitor and cuff carries different warranties in different countries
of purchase. Please see below for warranty length and contact information where you
purchased this product.
CONTACT INFORMATION
Distributed by:
A&D Engineering, Inc.
1756 Automation Parkway
San Jose, CA 95131
1-888-726-9966
www.andmedical.com
©2014 A&D Medical. All rights reserved. Specifications subject to change without notice.
I-MAN-UA-611-072016
Printed in China
In the United States of America:
5 year warranty
A&D Engineering, Inc.
1756 Automation Parkway
San Jose, CA 95131 U.S.A.
www.andmedical.com
1-888-726-9966
In Canada:
5 year warranty
Distributed in Canada by:
Auto Control Medical, Inc.
6695 Millcreek Drive, Unit 6
Mississauga, Ontario, L5N 5R8 Canada
www.lifesourcecanada.com
1-800-461-0991
In Latin America:
No warranty
Manufactured for:
A&D Medical
A division of A&D Engineering, Inc.
1756 Automation Parkway
San Jose, CA 95131
Toll-Free: 1-888-726-9966
www.andmedical.com
In Europe:
Various warranties - see packaging
All other countries:
Various warranties - see packaging
A&D Instruments LTD.
Unit 24/26 Blacklands Way,
Abingdon Business Park
Abingdon, Oxfordshire OX14 1DY
United Kingdom
www.aandd-eu.net
[44] (1235) 550420
A&D Company, Limited
1-243 Asahi, Kitamoto-shi,
Saitama, 364-8585 Japan
www.aandd.jp
[81] (48) 593-1119
Instruction Manual
ORIGINAL
1WMPD4002687B
UA-611
Symbols
Function/Meaning
Recommended Action
l
Artery position mark
Set the
l
mark on the artery of the upper arm
or in line with the ring finger on the inside of
the arm.
p
Index
––––––––––––
REF
Catalog number
––––––––––––
M
Proper fit range for the medium cuff. It's
printed on the cuff.
––––––––––––
L
Over range printed on the medium cuff and
AccuFit Plus cuff.
Use the large cuff instead of the medium cuff or
the AccuFit™ Plus cuff.
W
Proper fit range for the AccuFit™ Plus cuff.
It's printed on AccuFit™ Plus cuff.
––––––––––––
L
Proper fit range for the large cuff. It's
printed on the large cuff and AccuFit™
Plus cuff.
––––––––––––
S
Under range printed on the medium cuff and
AccuFit™ Plus cuff.
––––––––––––
M
Under range printed on the large cuff.
Use the medium cuff instead of the large cuff.
LOT
Lot number
––––––––––––