24
Press the UNITS key
to select an item
and
press the TARE key
to execute it.
To exit from the calibration mode, press the
ZERO
1
key or the CAL switch. The scale
returns to the weighing mode.
12.2. Gravity acceleration correction
When the scale is first used or has been moved to another location, it should be
calibrated using a calibration weight.
But if a calibration weight is not available, the gravity acceleration correction will
compensate the scale. Change the gravity acceleration value stored in the scale to the
value of the area where the scale will be used. Refer to the gravity acceleration map at
the end of this manual.
Note: Gravity acceleration correction is not required when the scale will be
calibrated using a calibration weight at the place where it is to be used.
1. Refer to “12.1. Calibration mode” to enter the calibration
mode. The gravity acceleration value is displayed.
2. Press the TARE key to enter the gravity acceleration
value setting mode.
3. Change the displayed value using the following keys.
UNITS To shift the blinking digit to the right.
PRINT To increase the value of the blinking digit
by one.
4. Press the TARE key. The setting value is stored and
“
end
”
is displayed.
5. When calibration using a calibration weight is to be
performed, go to step 3 of “12.3. Calibration using a
weight”.
To finish the setting procedure, press the ZERO
1
key or
the CAL switch. The scale returns to the weighing mode.
Press
Press
Press