29
Item
Parameter
Description
Serial interface
Response
aCk
0
No reply except data to commands
Only when the options,
RS-232C or RS-422/485,
are used.
1
Reply to commands
Comparator mode
Cp-l
0
Five-level
1
Three-level
2
Seven-level (portion weighing mode)
Comparison
conditions
Cp
0
Comparator disabled
Sets comparison
conditions
d = minimum display
in kg
1
Compares all data
2
Compares all stable data
3
Compares all data of
≥
+5d or
≤
-5d
4
Compares stable data of
≥
+5d or
≤
-5d
5
Compares data of
≥
+5d
6
Compares stable data of
≥
+5d
Key sensitivity
key
0
Low
1
Medium
2
High
Auto-tare
function
at
(See Note)
0
Auto-tare function disabled
Refer to “11. AUTO-
TARE FUNCTION”
1
Auto-tare function enabled
Auto-tare timing
at-t
(See Note)
0
Immediately after OK and stable
Timing to tare
automatically after
the comparison OK
and stable weight.
To be used with “
at 1
”.
1
0.5 second after OK and stable
2
1.0 second after OK and stable
3
1.5 seconds after OK and stable
4
2.0 seconds after OK and stable
5
2.5 seconds after OK and stable
6
3.0 seconds after OK and stable
7
4.0 seconds after OK and stable
8
5.0 seconds after OK and stable
9
6.0 seconds after OK and stable
Auto-tare for the
initial weight
at-f
(See Note)
0
Function disabled
Automatic operation
1
Tares the initial (container) weight.
Normal/Negative
comparison
Cp-p
0
Normal comparison
Refer to “11. AUTO-
TARE FUNCTION”
1
Negative comparison for take-away
check weighing
Factory setting
d = minimum display in kg, the minimum mass that can be weighed in kg
Even the counting mode uses “d” for judgment.
(
Note: In some countries or areas,
“
re5o
”, “
5t-b
”, “
at
”, “
at-t
” and “
at-f
”
are
not available for the Legal for Trade models.)