A&D SJ-1000H Instruction Manual Download Page 6

 5 

CALIBRATION

When to Calibrate 

Calibration may be required when it initially installed or it is moved a 
substantial distance. Also periodical calibration will be recommended. 

Calibration using a Calibration Weight 

1.  Enter calibration mode. 

 

Press the [ON/OFF] key to turn the power off. While 
pressing both the [RE-ZERO] and [UNITS] keys, 
press the [ON/OFF] key. 

 Cal  

  will be displayed. 

 

Or remove the calibration switch cover on the bottom 
and press the calibration switch while in the weighing 
mode. Then 

 Cal

  will be displayed. 

2. Zero calibration 

 

Press the [RE-ZERO] key to calibrate zero. 

 Cal0 

  will be displayed. 

 

Wait for the stable indicator to be displayed and press the [RE-ZERO] key.   

 Calf 

  will be displayed in a few seconds. If the zero calibration only is to 

be done, press the [UNITS] key. The scale will automatically return to the 
weighing mode.

3. Span calibration 

 When 

 Calf 

  is displayed, place the weight at the center of the platform. 

Wait for the stable indicator to be displayed, then press the [RE-ZERO] key. 
The display will show 

 end 

  and the scale will automatically return to the 

weighing mode. See “SPECIFICATIONS” about the calibration weight.

Calibration by Gravity Compensation 

The scale is calibrated at 9.798 m/s

2

 at the factory. If the acceleration of 

gravity at your location is not same and if you do not have calibration 
weights, you can calibrate the scale by compensating the acceleration of 
gravity. (Refer “The Value of Gravity at Various Locations”.) 

1.  Set a new acceleration value. 

 When 

 Cal  

  is displayed, press the [UNITS] key and 

 9.798

 will be 

displayed. Press the [RE-ZERO] key to increment the blinking digit and 
press the [UNITS] key to move the blinking digit. 

2.  Store the value in memory. 

 

While pressing the [UNITS] key, press and hold the [RE-ZERO] key and 
release the [UNITS] key. Then the display will show 

 end 

 and return to  

 Cal  

. Turn the scale off to finish the procedure. 

Calibration switch cover

Calibration switch

Summary of Contents for SJ-1000H

Page 1: ...SJ 1000H SJ 2000H SJ 5000H SJ 12KH WM PD4001280A ...

Page 2: ...facture When using the SJ H the following safety precautions should always be followed WARNING Use only specified AC adapter for the scale Other adapter may cause damage AC adapter is optional Internal service or adjustment to this product should be performed by a qualified person CAUTION Avoid installing the scale in direct sunlight which may cause discoloration or malfunctions Avoid over and or ...

Page 3: ...Tare function is used Polarity Battery indicator Indicates battery remaining amount It turns on about for 10 seconds when powered on Weighing unit k g lb or oz depends on model Weighing Platform ON OFF key Turns the scale power on off UNITS key Changes weighing units k g lb oz or lb oz RE ZERO key Sets display to zero or subtracts weight of container Spirit level Display Four leveling feet Battery...

Page 4: ...3 Insert new four batteries R20P LR20 D size into the battery box taking extreme care of the polarities Batteries are not included with shipment Connecting an AC adapter Optional The AC input requirement could be 100 120 or 230 volts 50 60Hz depending on the area where used so please verify that the adapter is correct Open the rubber cap and plug the AC adapter Setting up the Scale Place the scale...

Page 5: ...r Calibration may be required before weighing Read CALIBRATION first and if necessary calibrate your scale for accurate weighing Weighing Procedures 1 Turn on the scale Press the ON OFF key to turn on the scale Then the scale displays all segments for a few seconds and then 0 will be displayed 2 Start weighing If you do not use a container for weighing Verify the reading is 0 If not press the RE Z...

Page 6: ...When Calf is displayed place the weight at the center of the platform Wait for the stable indicator to be displayed then press the RE ZERO key The display will show end and the scale will automatically return to the weighing mode See SPECIFICATIONS about the calibration weight Calibration by Gravity Compensation The scale is calibrated at 9 798 m s 2 at the factory If the acceleration of gravity a...

Page 7: ...ht Approximately 1 9 kg 4 2 lb Calibration weight 1000g 0 1g 2000g 0 2g 5000g 0 5g 10 kg 1g Accessories This manual Options AC adapter The Value of Gravity at Various Locations Amsterdam 9 813 m s2 Athens 9 807 m s2 Auckland NZ 9 799 m s2 Bangkok 9 783 m s2 Birmingham 9 813 m s2 Brussels 9 811 m s2 Buenos Aires 9 797 m s2 Calcutta 9 788 m s2 Cape Town 9 796 m s2 Chicago 9 803 m s2 Copenhagen 9 815...

Page 8: ...ands Way Abingdon Business Park Abingdon Oxon OX14 1DY United Kingdom Telephone 44 1235 550420 Fax 44 1235 550485 German Scales Office Großbe Straße 13 b 22926 Ahrensburg GERMANY Telephone 49 0 4102 459230 Fax 49 0 4102 459231 A D MERCURY PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited 8th Floor Manhattan Bldg 36 2 Yoido ...

Reviews: