USER MENU DESCRIPTIONS

NAME/CODE

DESCRIPTION

CODE/VALUE

A1

Baud Rate

Selects the baud rate for data transmission through the serial port.

1200       

2400

√√√√

4800       9600

A2

Data Bits and
Parity

Selects the number of data bits and parity of serial transmission.

"8n" = 8 data bits with no parity bit and one stop bit

"7O" = 7 data bits with odd parity bit and one stop bit

"7E" = 7 data bits with even parity bit and one stop bit

"7n" = 7 data bits with no parity bit and two stop bits

8n

√√√√

7O
7E
7n

A3

Mode of Serial
Transmission

Selects when data will be sent out of the serial port to a printer or
computer:

"C" = Continuous mode; send data continuously

"d" = Demand mode; send data when a PRINT command is issued

from the printer, computer, or scale.

C

d

√√√√

A4

Serial Data
Format

Selects the data format to be transmitted via the serial port to a
printer or computer.

"0" = Consolidated Controls Format

"1" = Toledo 8213 Format          "2" = A&D/NCI/CAS Format

"3" = Transcell Technology Format

“4” = Detecto ASD Format        “5” = Triner Format

"6" = Fairbanks 70-2453-4 Format

0
1
2

3

√√√√

4
5
6

A5

Disable the
lb/kg Key

Allows the lb/kg key to be disabled so that an operator cannot acci-
dentally press the key and change the displayed units.
"0" = Disable the lb/kg key              "1" = Enable the lb/kg key

0

1

√√√√

Setup Menu Mode

SB-30/70/150 scale include a Setup menu which determines the scale’s display mode, allows
scale calibration and also contains the scale's adjustable functional parameters. It consists of 5
separate menu selections, each also with its own sub-menu of choices.

To change a listed parameter, you must first enter the Setup Menu mode. Once there, four of the
front panel keys become directional navigators to move around in the menus, and one key is used
to save or SET the selections.

Summary of Contents for SB-30

Page 1: ...S SB B S Se er ri ie es s D Di ig gi it ta al l B Be en nc ch h S Sc ca al le es s Instruction Manual Revision 1 3 April 5 2000 SB 30 SB 70 SB 150...

Page 2: ...150 lb 70 kg capaci ties It comes standard with a bright LED screen for easy readout a removable stain less steel platform and an RS 232C serial communication port These scales can be configured to d...

Page 3: ...e rocker switch to the ON 1 position The scale is now ready for use ON OFF SWITCH DC POWER JACK RS 232 PORT REMOTE DISPLAY ID TAG CALIBRATION SWITCH COVER SB 30 70 ONLY CONNECTOR SB 30 70 150 Back Pan...

Page 4: ...never the displayed weight is within 0 25 divisions of true zero NET Indicates that the scale is displaying net weight GROSS Indicates that the scale is displaying gross weight TARE Indicates that the...

Page 5: ...n NET GROSS Toggles between Gross and Net weight display if a tare has been established TARE Used to establish the weight on the platform as the tare weight You cannot establish zero or negative gross...

Page 6: ...wer off to the unit 2 While holding down the lb kg key turn the power back on 3 When the display shows A1 the unit is in User Menu mode and you can release the lb kg key Shown at right are the direc t...

Page 7: ...rial Data Format Selects the data format to be transmitted via the serial port to a printer or computer 0 Consolidated Controls Format 1 Toledo 8213 Format 2 A D NCI CAS Format 3 Transcell Technology...

Page 8: ...tional and SET key assign ments SET SETUP MENU CHART F1 Display Mode F6 Zero Range F9 Calib Unit 0 1 F11 Display Check F12 Calibration Press ZERO key to begin Press ZERO key to begin 0 1 100 1 9 To pl...

Page 9: ...ation Places scale into the calibration routine Scrolling down with the ZERO key one level begins the procedure Press ZERO key to begin sequence F13 Factory Reset This sub menu will reset all paramete...

Page 10: ...ged by pressing the PRINT key or the TARE key 5 After setting the exact value press the NET GROSS key to save the value 6 If the calibration was successful the display will show ECAL momentarily then...

Page 11: ...x 0 05 lb 0 70 x 0 02 kg CONSTRUCTION Base Steel and ABS Platform Stainless Steel Feet Non skid Hard Rubber DISPLAY 6 Digit 0 6 7 Segment Green LED OVER CAPACITY ANNUNCIATION 102 of Full Scale Capacit...

Page 12: ...lect the proper settings for the following using the scale s settings Baud Rate Data Bits Parity Stop Bits 1 4 Enable the CAPS LOCK key on your PC 5 This test verifies that the scale is transmitting p...

Page 13: ...ary Try re calibrating the scale if this doesn t solve the problem Otherwise possible load cell damage due to overloading LO Warmup Mode The scale has been turned on without the platter installed Inst...

Page 14: ...uch goods by Seller discloses that the nonconformity actually exists and was not caused by accident misuse neglect alteration improper installation improper or un authorized repair or improper testing...

Reviews: