12.Specifications - 12.2.Functional Specifications
146
1WMPD4004444
Playback processing Y-T waveform
FFT Analysis
X-Y waveform
Display position changes via swiping and pinch in/pinch out
operations to enlarge/reduce.
When the recorded data is NORMAL, FFT analysis function
processing is possible.
When the recorded data is NORMAL, X-Y processing is
possible,
sampling 1 kS/s or below.
*
1
Available when the optional remote control module is installed.
12.2.3. Memory Recording
Item
Specifications
Function
Records to the internal memory with high-speed sampling.
Recording device
Internal memory
Record blocks
(number of memory
divisions)
Points
2 GW
*
1
Divided into 1 to 200 user-defined blocks
The number of data items per channel that can be recorded
to a recording block
2 kW to 2 GW (selected in step 1-2-5)
Channels x points x blocks
≤
2 GW
Number of channels Analog
Logic
36 ch (max), 18 ch at 20 MS/s
144 ch (max)
Data format
NORMAL data
Sampling speed
20 MS/s to 10 S/min
The speed can be set to 1, 2, or 5 series
Information data
Records information related to the recorded data, such as the version of the product,
module configuration, channel settings, data format, and recording time.
Recorded data
Records input data and trigger information.
Playback processing Y-T waveform
Display position changes via swiping and pinch in/pinch out
operations to enlarge/reduce.
*
1
W(word) refers to a unit of data. 1 W = 2 bytes
Summary of Contents for RA3100 Omniace
Page 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Page 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Page 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...